Revolution CT, x-ray computed tomography, Model Numbers 5590000 and 5590000-6 Recall
Revolution CT, x-ray computed tomography, Model Numbers 5590000 and 5590000-6
FDA recall Class II
Ongoing
FDA lists this Class II recall for Revolution CT, x-ray computed tomography, Model Numbers 5590000 and 5590000-6. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2150-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 26, 2026
- FDA report date
- May 20, 2026
- Quantity in commerce
- 38 units
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
Revolution CT, x-ray computed tomography, Model Numbers 5590000 and 5590000-6
Affected model & UDI checker
Search the exact identifiers published for Z-2150-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 5590000 | Revolution CT, x-ray computed tomography, Model Numbers 5590000 and 5590000-6 | Check source notice | Not specified |
| Not specified | Revolution CT, x-ray computed tomography, Model Numbers 5590000 and 5590000-6 | Check source notice | 00840682118552 |
FDA code information
UDI-DI: 00840682118552; Serial Numbers:REVVX2300007CN 443327CN2 REVVX1600034CN REVVX2400004CN REVVX2200047CN 441716CN8 00000440713CN6 REVVX1800043CN 441839CN8 REVVX1700067CN REVVX1800007CN REVVX1800105CN REVVX1800048CN REVVX1900123CN REVVX1800046CN REVVX2000005CN REVVX2200009CN REVVX1600013CN REVVX1700101CN REVVX2200025CN REVVX2100028CN REVVX1900069CN REVVX2100020CN REVVX2100014CN REVVX2100050CN REVVX1600052CN 438902CN9 REVVX1900107CN 438369CN1 REVVX1900018CN REVVX1700071CN REVVX1700082CN 442792CN8 REVVX2200007CN REVVX1700104CN REVVX2000056CN REVVX1900086CN REVVX2200039CN
Why was this product recalled?
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did GE Medical Systems, LLC take?
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 3/26/26 was sent to customers. Actions to be taken by Customer/User Pending corrections by GE HealthCare, you can continue to use your system by following security and cybersecurity best practices. Before the interim solution (see Product Correction section below) is applied, please follow the instructions below as identified in the respective CT system User Manuals: CAUTION: 3D or slab reconstructions provide additional supplemental information, complementing diagnosis that should be based on classical techniques. WARNING: Do not use 3D or slab views only to perform any measurements (distance, angle, region of interest, report cursor, area, volume, etc.). Always check measurement points position and refer to 2D baseline views (acquisition images or reformatted images of minimal thickness) to confirm measurements. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via (https://gehealthcare-svc.my.site.com/publicForm/s/?formId=aGjUr000002jnD7) or print, fill out manually, scan and email to [email protected] Product Correction GE HealthCare will correct all affected products at no cost to you in two steps. A GE HealthCare representative will contact you to arrange for an interim solution to deactivate AW Server authentication through the CT system on the EHL-based Smart Subscription, and later, to install a CT software update. Note: With the interim solution, all alternative authentication mechanisms (direct application launch from CT console user interface, login using users defined in AW Server user database, federation with hospital user directory through LDAP) will remain available. Contact Information If you have any questions or concerns regarding this notification, please contact GE HealthCare
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2150-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026