Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2155-2026

Revolution Apex Select, X-ray Computed Tomography, Model Number 5590000-27 Recall

Revolution Apex Select, X-ray Computed Tomography, Model Number 5590000-27

FDA recall Class II

Ongoing

FDA lists this Class II recall for Revolution Apex Select, X-ray Computed Tomography, Model Number 5590000-27. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2155-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 26, 2026
FDA report date
May 20, 2026
Quantity in commerce
5 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

Revolution Apex Select, X-ray Computed Tomography, Model Number 5590000-27

Affected model & UDI checker

Search the exact identifiers published for Z-2155-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
5590000-27Revolution Apex Select, X-ray Computed Tomography, Model Number 5590000-27Check source noticeNot specified
Not specifiedRevolution Apex Select, X-ray Computed Tomography, Model Number 5590000-27Check source notice00195278469700
FDA code information

UDI-DI: 00195278469700; Serial Numbers: REV242400001CN REV242400004CN REV242300006CN REV242400002CN REV242300004CN

View the original FDA recall notice

Why was this product recalled?

GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did GE Medical Systems, LLC take?

An URGENT MEDICAL DEVICE CORRECTION notification letter dated 3/26/26 was sent to customers. Actions to be taken by Customer/User Pending corrections by GE HealthCare, you can continue to use your system by following security and cybersecurity best practices. Before the interim solution (see Product Correction section below) is applied, please follow the instructions below as identified in the respective CT system User Manuals: CAUTION: 3D or slab reconstructions provide additional supplemental information, complementing diagnosis that should be based on classical techniques. WARNING: Do not use 3D or slab views only to perform any measurements (distance, angle, region of interest, report cursor, area, volume, etc.). Always check measurement points position and refer to 2D baseline views (acquisition images or reformatted images of minimal thickness) to confirm measurements. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via (https://gehealthcare-svc.my.site.com/publicForm/s/?formId=aGjUr000002jnD7) or print, fill out manually, scan and email to [email protected] Product Correction GE HealthCare will correct all affected products at no cost to you in two steps. A GE HealthCare representative will contact you to arrange for an interim solution to deactivate AW Server authentication through the CT system on the EHL-based Smart Subscription, and later, to install a CT software update. Note: With the interim solution, all alternative authentication mechanisms (direct application launch from CT console user interface, login using users defined in AW Server user database, federation with hospital user directory through LDAP) will remain available. Contact Information If you have any questions or concerns regarding this notification, please contact GE HealthCare

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology