Bridge to Life EasiSlush, (Sodium Chloride Solution for Sterile Slush Preparation) (0.9% Sodium Chloride Irrigation, USP), Part Number REF BTLE-1250 Recall

Bridge to Life EasiSlush, (Sodium Chloride Solution for Sterile Slush Preparation) (0.9% Sodium Chloride Irrigation, USP), Part Number REF BTLE-1250

FDA recall Class II

Ongoing

FDA lists this Class II recall for Bridge to Life EasiSlush, (Sodium Chloride Solution for Sterile Slush Preparation) (0.9% Sodium Chloride Irrigation, USP), Part Number REF BTLE-1250. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2157-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 12, 2025
FDA report date
Jul 30, 2025
Quantity in commerce
4528 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Bridge to Life EasiSlush, (Sodium Chloride Solution for Sterile Slush Preparation) (0.9% Sodium Chloride Irrigation, USP), Part Number REF BTLE-1250

Affected model & UDI checker

Search the exact identifiers published for Z-2157-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
BTLE-1250Bridge to Life EasiSlush, (Sodium Chloride Solution for Sterile Slush Preparation) (0.9% Sodium Chloride Irrigation, USP), Part Number REF BTLE-1250Check source noticeNot specified
FDA code information

UDI 10860001991205-240206-260206-240372708 (case), 00860001991208-240206-260206-240372708 (bag), Lot Number 240372708; UDI 10860001991205-240212-260212-240402836 (case), 00860001991208-240212-260212-240402836 (bag), Lot Number 240402836; UDI 10860001991205-240214-260214-240402837 (case), 00860001991208-240214-260214-240402837 (bag), Lot Number 240402837.

View the original FDA recall notice

Why was this product recalled?

The firm issued a field safety notice after becoming aware of three lots of products not having quarterly dose audits being completed on time in the first quarter of 2024. Routine lot by lot testing was completed properly however the quarterly dose audit (sampling/testing to monitor the gamma sterilization process and confirm the validated parameters remain effective to assure sterility and bioburden reduction per requirements) was not completed on time.

What actions did Bridge to Life Ltd take?

Bridge to Life issued an IMPORTANT FIELD SAFETY NOTICE to its consignees on 06/12/2025 via email. The notice explained the issue and requested the following: Actions required by the customer: 1. To be aware and acknowledge this field safety notice. 2. We are requesting customers to conduct a review of the existing EasiSlush inventory to determine if you have any quantities of the following three lot numbers 240372708, 240372836, 240402837. 3. If you may have further distributed the product, we request that quantity be included in your review and please advise us on where your product is forwarded with contact information. 4. Immediately notify Bridge to Life of the quantities in inventory. Fill out FSN Customer Reply Form below and send it back to the email address [email protected] 5. FSN Customer Reply Form must be returned no later than June 18, 2025. 6. Lastly, please confirm no adverse events related to the subject lots and advise if you have any additional questions or concerns. For questions, please contact [email protected] or 800-667-9524.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.