Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS Recall
Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS
FDA recall Class I
Ongoing
FDA lists this Class I recall for Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2173-2025
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Jul 14, 2025
- FDA report date
- Aug 20, 2025
- Quantity in commerce
- 36705 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS
Affected model & UDI checker
Search the exact identifiers published for Z-2173-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 40700BAXUS | Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS | Check source notice | Not specified |
| Not specified | Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS | Check source notice | 05413765851797 |
FDA code information
UDI/DI 05413765851797, All Serial Numbers
Why was this product recalled?
Baxter is issuing an Urgent Medical Device Correction for the Novum IQ large volume pump (LVP) due to the potential for underinfusion when transitioning from a flow rate to a higher flow rate that is more than double. (e.g., rate change or bolus). The level of underinfusion is variable based on the current infusion rate, the duration the pump has been running at this flow rate, and the magnitude of the rate change. The longer the duration the pump has been running at the current infusion rate and the larger the magnitude of the rate change, the larger the underinfusion that would be experienced. Additionally, Baxter has identified an increase in customer reports of over and underinfusion potentially due to set misloading. Failure to properly load the tubing into the pump channel may result in the pump infusing at a rate higher or lower than programmed. Consistent with the instructions for use, customers should ensure that: 1) The door is fully open before loading the set. 2) The tubing is taut and loaded without slack in the pumping channel.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Baxter Healthcare Corporation take?
Baxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 07/14/2025 via USPS First Class Mail. The notice explained the potential for under-infusion due to 1) when transitioning from a lower to higher flow rate, and/or 2) set misloading; hazard involved and requested the following: Actions to be Taken by Customers 1. Users should rely on their clinical judgment. When the delay associated with changing the pump and infusion set would NOT prove detrimental to the patient, users should change out the pump and infusion set prior to initiating a bolus infusion or a rate change of greater than 100%. 2. If the delay associated with changing the pump and/or infusion set would prove unacceptable, prior to initiating a bolus infusion or a rate change of greater than 100%, users should move the administration set downstream (towards the patient) approximately 0.5 inches by following steps A through C below. A. Please follow the instructions for unloading the set in section 4.4, Unloading an Administration Set, of the Novum IQ LVP Operator s Manual (refer to enclosed Attachment A). - To prevent free flow, ensure downstream roller clamp is fully closed. B. Once unloaded, move the administration set downstream by moving the slide clamp toward the container by 0.5 inches. C. After moving the slide clamp, reload the administration set. Please follow the instructions for loading the set in section 4.3, Loading an Administration Set, of the Novum IQ LVP Operator s Manual (refer to enclosed Attachment A). - Once the tubing has been loaded, the door closed and the slide clamp has been ejected, ensure that the downstream roller clamp is fully open. D. Initiate replacement of the administration set at the first safe opportunity to do so, or check regularly to ensure the infusion proceeds at the expected rate. 3. To prevent misloading, please follow the instructions for loading the set in section 4.3, Loading an Administration Set, of the Novum
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2173-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology