Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an accessory to the Tempus Pro Monitor, REF: 00-1007-R, 00-1024-R, 00-1026-R Recall
Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an accessory to the Tempus Pro Monitor, REF: 00-1007-R, 00-1024-R, 00-1026-R
FDA recall Class II
Ongoing
FDA lists this Class II recall for Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an accessory to the Tempus Pro Monitor, REF: 00-1007-R, 00-1024-R, 00-1026-R. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2175-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 30, 2025
- FDA report date
- Aug 6, 2025
- Quantity in commerce
- 55
- Distribution
- See key details
- FDA cause
- Software change control
What is being recalled?
Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an accessory to the Tempus Pro Monitor, REF: 00-1007-R, 00-1024-R, 00-1026-R
Affected model & UDI checker
Search the exact identifiers published for Z-2175-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 00-1007-R | Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an accessory to the Tempus Pro Monitor, REF: 00-1007-R, 00-1024-R, 00-1026-R | Check source notice | Not specified |
| 00-1024-R | Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an accessory to the Tempus Pro Monitor, REF: 00-1007-R, 00-1024-R, 00-1026-R | Check source notice | Not specified |
| 00-1026-R | Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an accessory to the Tempus Pro Monitor, REF: 00-1007-R, 00-1024-R, 00-1026-R | Check source notice | Not specified |
FDA code information
REF/UDI-DI: 00-1007-R/05060472440013, 05060472442918; 00-1024-R/05060472441027, 05060472442925; 00-1026-R/05060472441058, 05060472442932. Software Version: vx.28
Why was this product recalled?
Kit USB flash drive contains outdated software, which could result in reverting patient monitor to outdated software and reintroducing the video laryngoscope issue (related to an August 2023 recall), which could lead to delay in diagnosis, delayed treatment, hypoxia due to unexpected loss of video laryngoscopy and all other monitor measurements during system restart.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Remote Diagnostic Technologies Ltd. take?
On 5/30/3035, Philips mailed field safety notices to customers and distributors who were informed of the following: Reverting to the outdated software version reintroduces the video laryngoscope issue previously corrected in FSN-2023-CC-EC-005 (Unexpected Device Error When Used with Tempus Pro Video Laryngoscope). Customers were asked to do the following: 1) Dispose of USB Flash Drive. 2) Ensure monitor is on Software version vx.34 or higher. 4) If your device(s) is on a software version below vx.34, contact firm representative for software update instructions. 5) Complete and return response form via email to [email protected] In addition, distributors were asked to do the following: 1) Modify the correction response form and substitute your firm's email and fax information. Send a copy of the correction notice and modified response form to each customer to whom you distribute the affected product. If you need any further information or support contact firm at (800) 722-9377 or email [email protected]
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2175-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology