Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2179-2025

BD Alaris Pump Module Model, 8100 with Guardrails, Suite MX software versions Recall

BD Alaris Pump Module Model, 8100 with Guardrails, Suite MX software versions

FDA recall Class I

Ongoing

FDA lists this Class I recall for BD Alaris Pump Module Model, 8100 with Guardrails, Suite MX software versions. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2179-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Jul 8, 2025
FDA report date
Aug 13, 2025
Quantity in commerce
33,162,680 units total
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

BD Alaris Pump Module Model, 8100 with Guardrails, Suite MX software versions

Affected model & UDI checker

Search the exact identifiers published for Z-2179-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
8100-BD Alaris Pump Module Model, 8100 with Guardrails, Suite MX software versionsCheck source noticeNot specified
FDA code information

BD Alaris" Pump Module Model 8100- All Serial Numbers/ UDI(s): 10885403222054, 10885403517723, 10885403810015, 10885403810039, 10885403810046

View the original FDA recall notice

Why was this product recalled?

Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.

What actions did CareFusion 303, Inc. take?

On December 12, 2025, CareFusion 303, Inc. (BD) issued a updated "Urgent Medical device Recall Notice via E-Mail and physical mail. The updated notice includes important information on infusion sets as well as labeling updates. Previous notifications were sent on July 8,2025, July 17, 2025, September 11, 2025 and September 12, 2025. CareFusion ask consignees to take the following actions: 1. Review your inventory for the catalog numbers included in the updated letter. Remove and discard any remaining inventory per your facility guidelines. 2.To receive credit or replacement product, please return the completed customer response form attached to this notice. 3.Please contact your BD Clinical Consultants and Account Executives for alternative sets information. 4. Review the updated sets performance data included in the letter. 5. Circulate this notice within your facility network to ensure that all concerned personnel are made aware of this issue. 6. Distributors should provide a copy of this notice to all customers who may have purchased an Alaris Pump Module. If you require further assistance, please contact: BD Technical Support at Phone: 888-812-3229; Phone hours: 6:00am to 5:00pm PT Monday Friday and Email: [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology