BritePro Solo Single-Use Fiber Optic Laryngoscope Handle Refs:040-310AU and 040-310U Recall
BritePro Solo Single-Use Fiber Optic Laryngoscope Handle Refs:040-310AU and 040-310U
FDA recall Class I
Ongoing
FDA lists this Class I recall for BritePro Solo Single-Use Fiber Optic Laryngoscope Handle Refs:040-310AU and 040-310U. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2187-2025
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Jul 11, 2025
- FDA report date
- Aug 13, 2025
- Quantity in commerce
- N/A
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
BritePro Solo Single-Use Fiber Optic Laryngoscope Handle Refs:040-310AU and 040-310U
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
Lots: 2007xxxxx to 2012xxxxx (July-2020 to Dec-2020)
Why was this product recalled?
Laryngoscope handles may not illuminate as intended.
What actions did Flexicare Medical (Dongguan) Ltd. take?
On July 11, 2025, Flexicare Medical Limited issued a Urgent: Medical Device Recall notification to affected consignees via E-Mail. Flexicare ask consignees to take the following actions: 1. Review Inventory for affected product and immediately quarantine. 2. All affected devices must be manually checked using the following procedure within customer notification. 3. All devices which demonstrate a failure to illuminate shall be withdrawn from use, quarantined, and returned. 4. If tested devices illuminate, they can continue to be used. Please ensure you subsequently continue to follow the Pre-Use Checks information as per the IFU. 5. Share this notice with any users of the product within your distribution network. 6. If your firm has further distributed the product, please provide them with the notification and inform them to notify their customers. 7. Adverse reactions or quality problems experienced with the use of this product may be reported to FDA's MedWatch Adverse Event Reporting program . 8. In the event of light failure, the caregiver should revert to the local standard of care, including oxygenation, sourcing a functional laryngoscope, or relying on another method to provide patient ventilation.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2187-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.