Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35 Recall
Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35
FDA recall Class II
Ongoing
FDA lists this Class II recall for Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2192-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Mar 21, 2025
- FDA report date
- Aug 6, 2025
- Quantity in commerce
- 1,064
- Distribution
- See key details
- FDA cause
- Process change control
What is being recalled?
Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35
Affected model & UDI checker
Search the exact identifiers published for Z-2192-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 2.70S | Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35 | Check source notice | Not specified |
| 3.20L | Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35 | Check source notice | Not specified |
| 4.50 | Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35 | Check source notice | Not specified |
| 5.30 | Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35 | Check source notice | Not specified |
| 6.35 | Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35 | Check source notice | Not specified |
FDA code information
UDI-DI: B775270S0, B775320L0, B7754500, B7755300, B7756350. Lot: 241002450, 241002530, 241009450, 241009530, 241016450, 241016530, 241023450, 241023530, 241030450, 241030530, 24112027S, 24121827S, 24122727S, 25012227S, 24082832L, 24102332L, 241227450, 241220635, 250122635, 241227530, 250122530.
Why was this product recalled?
Drill for use in neurosurgical procedures may have a reverse battery orientation, which may increase surgical time.
What actions did PHASOR HEALTH, LLC take?
Starting on 3/21/2025, device correction and removal began and a correction and removal notice was emailed. Customers were asked to do the following: 1) Review inventory (without opening the sterile transparent barrier) to determine if any of the affected lots remain in inventory. If yes, remove those lots from inventory. Do not use drills with reverse battery orientation. 2) If unable to review inventory soon, the Company will come to your site within 30 business days of this letter and inspect your inventory (within 20 minutes total) and replace any defective drills with new verified stock (same catalog number). 3) Contact firm at [email protected] to report results of inventory inspection and to facilitate device return. 4) Complete and return the acknowledgement and receipt form via email to [email protected] Please contact Firm at [email protected] or (832) 982-1234 for more information or questions
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2192-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology