Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2199-2025

SwabFlush Prefilled Syringe with ICU Medical SwabCaps, 10mL Model no. EMZE010301 Recall

SwabFlush Prefilled Syringe with ICU Medical SwabCaps, 10mL Model no. EMZE010301

FDA recall Class II

Ongoing

FDA lists this Class II recall for SwabFlush Prefilled Syringe with ICU Medical SwabCaps, 10mL Model no. EMZE010301. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2199-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 16, 2025
FDA report date
Aug 6, 2025
Quantity in commerce
1,427,100 eaches
Distribution
See key details
FDA cause
Process control

What is being recalled?

SwabFlush Prefilled Syringe with ICU Medical SwabCaps, 10mL Model no. EMZE010301

Affected model & UDI checker

Search the exact identifiers published for Z-2199-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
EMZE010301SwabFlush Prefilled Syringe with ICU Medical SwabCaps, 10mL Model no. EMZE010301Check source noticeNot specified
Not specifiedSwabFlush Prefilled Syringe with ICU Medical SwabCaps, 10mL Model no. EMZE010301Check source notice00363807103016
Not specifiedSwabFlush Prefilled Syringe with ICU Medical SwabCaps, 10mL Model no. EMZE010301Check source notice10363807103013
Not specifiedSwabFlush Prefilled Syringe with ICU Medical SwabCaps, 10mL Model no. EMZE010301Check source notice20363807103010
FDA code information

UDI-DI 00363807103016 (ea) 10363807103013 (box) 20363807103010 (case) Lots 3144609 3144857 3144886 3144967 3145090 3145186

View the original FDA recall notice

Why was this product recalled?

SwabFlush syringes are affected by the ICU Medical recall of their SwabCap Disinfecting Cap due to a manufacturing issue which affects the seal between the foil lid and plastic container of the Swab Cap. The issue may result in a reduced level of disinfection of the luer access valve, potentially resulting in patient infection/sepsis.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did MEDLINE INDUSTRIES, LP - Northfield take?

On June 16, 2025, the firm notified customers of the affected product. Customers were instructed to destroy the affected product for credit.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology