FDA RECALL RECORD · Z-2213-2025

MR 5300 Model Numbers (REF): (1) 782110, (2) 782135 (China ONLY), (3) 782152; Recall

MR 5300 Model Numbers (REF): (1) 782110, (2) 782135 (China ONLY), (3) 782152;

FDA recall Class II

Ongoing

FDA lists this Class II recall for MR 5300 Model Numbers (REF): (1) 782110, (2) 782135 (China ONLY), (3) 782152;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2213-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 30, 2025
FDA report date
Aug 6, 2025
Quantity in commerce
551 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

MR 5300 Model Numbers (REF): (1) 782110, (2) 782135 (China ONLY), (3) 782152;

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Model Numbers (REF): (1) 782110, (2) 782135, (3) 782152; UDI-DI: (1) 00884838099364, (2) None, (3) 00884838112841; Serial Numbers: (1) 63000, 63001, 63006, 63007, 63008, 63009, 63010, 63011, 63012, 63013, 63014, 63015, 63016, 63018, 63019, 63021, 63022, 63023, 63024, 63025, 63026, 63027, 63028, 63029, 63030, 63031, 63032, 63033, 63034, 63035, 63036, 63038, 63039, 63041, 63042, 63043, 63044, 63046, 63047, 63049, 63050, 63051, 63053, 63054, 63057, 63058, 63059, 63060, 63062, 63063, 63064, 63066, 63067, 63068, 63069, 63070, 63071, 63072, 63074, 63075, 63076, 63078, 63079, 63080, 63082, 63083, 63084, 63085, 63086, 63087, 63090, 63091, 63092, 63093, 63094, 63095, 63096, 63097, 63098, 63099, 63100, 63103, 63104, 63105, 63106, 63107, 63108, 63109, 63110, 63111, 63112, 63113, 63115, 63116, 63119, 63121, 63122, 63123, 63125, 63126, 63127, 63128, 63130, 63132, 63134, 63135, 63136, 63137, 63138, 63139, 63140, 63141, 63142, 63143, 63144, 63146, 63148, 63149, 63150, 63151, 63152, 63153, 63154, 63155, 63156, 63157, 63159, 63160, 63161, 63162, 63164, 63165, 63166, 63168, 63169, 63172, 63174, 63175, 63176, 63177, 63178, 63182, 63183, 63186, 63187, 63188, 63190, 63191, 63193, 63194, 63195, 63196, 63197, 63198, 63199, 63200, 63201, 63202, 63203, 63210, 63211, 63218, 63219, 63220, 63221, 63222, 63223, 63224, 63226, 63227, 63228, 63229, 63235, 63236, 63237, 63238, 63239, 63240, 63243, 63244, 63247, 63248, 63250, 63252, 63253, 63257, 63258, 63259, 63260, 63261, 63262, 63263, 63264, 63265, 63266, 63267, 63268, 63269, 63270, 63271, 63272, 63273, 63274, 63275, 63276, 63277, 63281, 63282, 63283, 63284, 63285, 63286, 63287, 63288, 63289, 63290, 63292, 63294, 63295, 63296, 63297, 63299, 63300, 63301, 63302, 63303, 63304, 63305, 63306, 63307, 63308, 63309, 63311, 63312, 63313, 63314, 63315, 63316, 63318, 63319, 63322, 63323, 63325, 63328, 63330, 63331, 63333, 63334, 63336, 63337, 63340, 63343, 63344, 63346, 63352, 63355, 63356, 63357, 63360, 63361, 63363, 63364, 63365, 63366, 63367, 63369, 63370, 63371, 63372, 63374, 63375, 63379, 63380, 63383, 63384, 63385, 63386, 63388, 63389, 63393, 63394, 63396, 63397, 63398, 63400, 63401, 63402, 63408, 63409, 63410, 63411, 63412, 63413, 63416, 63418, 63419, 63422, 63424, 63425, 63428, 63429, 63430, 63431, 63432, 63433, 63434, 63435, 63437, 63438, 63439, 63440, 63442, 63443, 63445, 63446, 63449, 63451, 63452, 63453, 63455, 63456, 63457, 63461, 63464, 63466, 63474, 63475, 63477, 63478, 63482, 63484, 63487, 63488, 63489, 63492, 63493, 63494, 63497, 63498, 63499, 63500, 63502, 63503, 63504, 63506, 63507, 63510, 63513, 63514, 63515, 63519, 63520, 63521, 63522, 63523, 63524, 63530, 63531, 63532, 63533, 63534, 63541, 63547, 63548, 63550, 63555, 63557, 63559, 63560, 63561, 63562, 63563, 63564, 63565, 63566, 63567, 63569, 63570, 63571, 63572, 63576, 63583, 63586, 63587, 63589, 63593, 63596, 63597, 63598, 63599, 63603, 63604, 63609, 63610, 63612, 63614, 63615, 63616, 63618, 63623, 63624, 63626, 63630, 63631, 63632, 63633, 63639, 63650, 63664, 63665, 63666, 63668, 63672, 63673, 63674, 63679, 63684, 63694, 63696, 63697, 63698, 63699, 63704, (2) 28001, 28002, 28003, 28004, 28006, 28008, 28009, 28010, 28011, 28012, 28013, 28014, 28015, 28017, 28021, (3) 69000, 69003, 69004, 69005, 69006, 69007, 69008, 69010, 69011, 69012, 69014, 69015, 69016, 69017, 69019, 69020, 69021, 69022, 69023, 69024, 69026, 69027, 69029, 69030, 69031, 69032, 69034, 69036, 69037, 69041, 69043, 69044, 69045, 69046, 69048, 69051, 69052, 69053, 69055, 69056, 69058, 69060, 69061, 69063, 69064, 69065, 69066, 69067, 69068, 69069, 69070, 69071, 69072, 69073, 69075, 69076, 69077, 69078, 69080, 69081, 69082, 69085, 69086, 69087, 69089, 69092, 69093, 69094, 69095, 69096, 69097, 69098, 69099, 69100, 69102, 69103, 69106, 69107, 69110, 69111, 69112, 69115, 69116, 69121, 69122, 69123, 69124, 69126, 69127, 69128, 69130, 69131, 69133, 69136, 69137, 69141, 69142, 69145, 69147, 69150, 69154, 69161, 69169, 69170, 69171, 69172, 69175, 69176, 69178, 69179, 69195, 69216, 69233;

View the original FDA recall notice

Why was this product recalled?

The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Philips North America take?

On June 30, 2025, URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. 4. Actions that should be taken by the customer / user in order to prevent risks for patients or users 4.1. When using any affected system identified according to the information in section 3, follow the instructions below. 4.2. Pay particular attention if a scan interruption occurs and a scan abort symbol is encountered, which may appear in the user interface (UI) with the symbols shown in Figure 4 below: Figure 4. Scan Abort Symbols 4.3. When scan aborts occur the symbol is shown and applicable messages are presented in the message box (see Figure 5 and Figure 6 examples) 4.3.1. Check the status of the error messages in the User Interface (UI) screen. Click to expand the message box in case previous messages are hidden, see orange box in Figure 5 and Figure 6 below. If possible, correct the issue that is relevant to the error shown and continue scanning, for example: RF door must be closed while performing scan. Scan aborted. Patient support is moved while scanning, scan stopped. Coil A (or B) failure, coil disconnected? . 4.3.2. If, after a scan abort, one of the following conditions occur five times in a row stop scanning immediately and contact Philips Service to describe the problem: If there is no error message in the message box If the error message meaning is not clear 4.3.3. If the error message Gradient amplifier Rack Fault is received two times in a row, or if the error message Gradient amplifier Rack Fault is preceded or followed by an aborted scan with no message, then stop scanning immediately and contact Philips Service to describe the problem. 4.3.4. If you encounter any of these conditions described in 4.3.2 or 4.3.3 do not attempt any additional/further scans, including without limitation do not attempt any phantom scanning until your system has been checked and released by a Philips Service representative. 4.4. If smoke and/or fire is detected

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.