Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2230-2025

RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097. Recall

RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.

FDA recall Class II

Ongoing

FDA lists this Class II recall for RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2230-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 26, 2025
FDA report date
Aug 6, 2025
Quantity in commerce
1,477 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.

Affected model & UDI checker

Search the exact identifiers published for Z-2230-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedRAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.Check source notice00630414482163
FDA code information

Material Number: 10329097. GTIN Number: 00630414482163. Lot Number: WW/08925. Expiration Date: 12/03/2025.

View the original FDA recall notice

Why was this product recalled?

Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Siemens Healthcare Diagnostics, Inc. take?

Siemens Healthcare sent an Urgent Medical Device Correction (UMDC) letter to consignees (via FedEx in the US) on around 06/26/2025. The communication letter informs consignees of the issue and provided actions for consignees to take. Consignees were instructed to review the letter with the Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable, replace the both the measurement cartridge and the wash/waste cartridge if still using a wash/waste cartridge from lot WW/08925 with RAPIDPoint systems, discontinue use of the wash/waste cartridges from lot WW/08925 and dispose any remaining inventory in possession. No-cost replacement wash/waste cartridges should be made through the local Siemens Healthineers or distributor office. Consignees were also instructed to complete and return the Field Correction Effectiveness Check Form included with letter, notify customers if products were further distributed, and to retain this letter with your laboratory records and forward this letter to those who may have received this product at your site.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.