Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2246-2025

GEN LAPAROSCOPY PACK. Medical convenience kit. Recall

GEN LAPAROSCOPY PACK. Medical convenience kit.

FDA recall Class II

Ongoing

FDA lists this Class II recall for GEN LAPAROSCOPY PACK. Medical convenience kit.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2246-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 11, 2025
FDA report date
Aug 13, 2025
Quantity in commerce
321 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

GEN LAPAROSCOPY PACK. Medical convenience kit.

Affected model & UDI checker

Search the exact identifiers published for Z-2246-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
VMCC007-12GEN LAPAROSCOPY PACK. Medical convenience kit.Check source noticeNot specified
Not specifiedGEN LAPAROSCOPY PACK. Medical convenience kit.Check source notice10809160357200
FDA code information

Model No. VMCC007-12; UDI: 10809160357200; Kit Lot No. 1592617, 1598707, 1608042, 1608043, 1621654.

View the original FDA recall notice

Why was this product recalled?

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

What actions did AVID Medical, Inc. take?

Consignees were sent an URGENT: MEDICAL DEVICE FIELD CORRECTION notice, dated 6/11/25, via email. The notification instructs consignees to not use the CLEARIFY Visualization System (component #5303907) and discard. All other components in the kit may be continued to be used. Consignees are to add warning labels to affected procedure trays to ensure the affected component is discarded. The provided Recall Response Form is to be completed and returned to Owens&Minor. Any quality problems or adverse events associated with recalled devices may be reported via email to [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology