Brand Name: T4 Total ELISA Product Name: T4 Total ELISA Recall
Brand Name: T4 Total ELISA Product Name: T4 Total ELISA Model/Catalog Number: EIA-4568 Product Description: T4 Total ELISA
FDA recall Class III
Ongoing
FDA lists this Class III recall for Brand Name: T4 Total ELISA Product Name: T4 Total ELISA. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2265-2025
- Recall classification
- Class III
- Enforcement status
- Ongoing
- Initiation date
- Jun 4, 2025
- FDA report date
- Aug 13, 2025
- Quantity in commerce
- 47 units
- Distribution
- See key details
- FDA cause
- Reprocessing Controls
What is being recalled?
Brand Name: T4 Total ELISA Product Name: T4 Total ELISA Model/Catalog Number: EIA-4568 Product Description: T4 Total ELISA
Affected model & UDI checker
Search the exact identifiers published for Z-2265-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| EIA-4568 | Brand Name: T4 Total ELISA Product Name: T4 Total ELISA | Check source notice | Not specified |
| Not specified | Brand Name: T4 Total ELISA Product Name: T4 Total ELISA | Check source notice | 04048474045680 |
FDA code information
Model/Catalog Number: EIA-4568; UDI-DI: 04048474045680; Lot number: 304K034;
Why was this product recalled?
An incoming complaint reported that the expiration date on the kit box label and Certificate of Analysis (CoA) exceeded the shelf life of the standards and control solutions in the kit by 2 months (i.e., 2026-03-31 instead of 2026-01-31). The expiration dates on the bottle labels were correct (i.e., 2026-01-31).
What actions did DRG International, Inc. take?
On June 4, 2025 FIELD SAFETY NOTICE letters were sent to customers. Action to be taken by distributors and end-users: Please check your inventory for the respective lot. Please use the CoA provided with this FSN. If you require corrected outer kit box label for your documentation, please contact DRG. If you have transferred any of the affected products to another laboratory, please provide them with a copy of this letter. If you have any questions concerning this issue, please do not hesitate to contact us.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2265-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026