Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2275-2025

One-piece Guedel airway, size 2, ISO 8.0, green. Recall

One-piece Guedel airway, size 2, ISO 8.0, green. Model Number: 1112080.

FDA recall Class II

Ongoing

FDA lists this Class II recall for One-piece Guedel airway, size 2, ISO 8.0, green.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2275-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 20, 2025
FDA report date
Aug 13, 2025
Quantity in commerce
2,800 pieces
Distribution
See key details
FDA cause
Process control

What is being recalled?

One-piece Guedel airway, size 2, ISO 8.0, green. Model Number: 1112080.

Affected model & UDI checker

Search the exact identifiers published for Z-2275-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
1112080One-piece Guedel airway, size 2, ISO 8.0, green.Check source noticeNot specified
FDA code information

Model Number: 1112080. UDI-DI (Case): 05030267040698, UDI-DI (Single Unit): 5030267050659. Lot Number: 32414447

View the original FDA recall notice

Why was this product recalled?

Potential contamination with small burrs, which if detach could be inhaled and result in potential complications such as airway obstruction, tissue irritation, inflammation and infection.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Intersurgical Inc take?

Intersurgical notified consignees on 06/20/2025 via email with recall letter. Consignees were instructed to immediately discontinue use and quarantine any affected units, immediately notify customers if product has been further distributed, arrange for the return of affected units and complete the Customer Product Recall Acknowledgement Form.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history