One-piece Guedel airway, size 3, ISO 9.0, yellow. Recall
One-piece Guedel airway, size 3, ISO 9.0, yellow. Model Number: 1113090.
FDA recall Class II
Ongoing
FDA lists this Class II recall for One-piece Guedel airway, size 3, ISO 9.0, yellow.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2276-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jun 20, 2025
- FDA report date
- Aug 13, 2025
- Quantity in commerce
- 8,640 pieces
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
One-piece Guedel airway, size 3, ISO 9.0, yellow. Model Number: 1113090.
Affected model & UDI checker
Search the exact identifiers published for Z-2276-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 1113090 | One-piece Guedel airway, size 3, ISO 9.0, yellow. | Check source notice | Not specified |
FDA code information
Model Number: 1113090. UDI-DI (Case): 05030267153961, UDI-DI (Single Unit): 5030267050680. Lot Number: 32420657
Why was this product recalled?
Potential contamination with small burrs, which if detach could be inhaled and result in potential complications such as airway obstruction, tissue irritation, inflammation and infection.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Intersurgical Inc take?
Intersurgical notified consignees on 06/20/2025 via email with recall letter. Consignees were instructed to immediately discontinue use and quarantine any affected units, immediately notify customers if product has been further distributed, arrange for the return of affected units and complete the Customer Product Recall Acknowledgement Form.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2276-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology