Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2278-2025

FARAWAVE 1.0 Pulsed Field Ablation Catheter, Product ID M004PF41M401 Recall

FARAWAVE 1.0 Pulsed Field Ablation Catheter, Product ID M004PF41M401

FDA recall Class II

Ongoing

FDA lists this Class II recall for FARAWAVE 1.0 Pulsed Field Ablation Catheter, Product ID M004PF41M401. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2278-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 25, 2025
FDA report date
Aug 20, 2025
Quantity in commerce
445 units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

FARAWAVE 1.0 Pulsed Field Ablation Catheter, Product ID M004PF41M401

Affected model & UDI checker

Search the exact identifiers published for Z-2278-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedFARAWAVE 1.0 Pulsed Field Ablation Catheter, Product ID M004PF41M401Check source notice00191506043148
FDA code information

UDI-DI: 00191506043148; Batch Numbers: 36543581 36543644 36543646 36568553 36572565 36572568 36572571 36572576 36578308 36578314 36579110 36598352 36598809 36599402 36600406 36600543 36600544 36600545 36600578

View the original FDA recall notice

Why was this product recalled?

The potentially impacted units were manufactured using specific equipment that may have caused cracks in the electrode bands on the catheter's distal end.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did BOSTON SCIENTIFIC CORPORATION take?

An Urgent Medical Device Removal - Immediate Action Required notification letter dated June 2025 was sent to customers. Actions 1. Do NOT use affected product (Attachment 1), remove affected devices from your facility s inventory, segregate the units in a secure place until they can be returned to Boston Scientific. 2. Immediately post this information in a visible location near the affected products to ensure this information is readily accessible to all handlers and users of the device. 3. Forward this notice to any healthcare professional from your organization for awareness and to any organization where affected devices have been transferred. 4. Complete and return the enclosed Reply Verification Tracking Form per the enclosed instructions on page five. 5. Return affected product. Email: [email protected] Reach out to your local Boston Scientific representative with any questions.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology