Covidien EndoStitch" Polysorb Single Use Loading Unit; Product numbers: 170053 and 170071; Recall
Covidien EndoStitch" Polysorb Single Use Loading Unit; Product numbers: 170053 and 170071;
FDA recall Class II
Ongoing
FDA lists this Class II recall for Covidien EndoStitch" Polysorb Single Use Loading Unit; Product numbers: 170053 and 170071;. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2281-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 15, 2026
- FDA report date
- Jun 10, 2026
- Quantity in commerce
- 2448 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Covidien EndoStitch" Polysorb Single Use Loading Unit; Product numbers: 170053 and 170071;
Affected model & UDI checker
Search the exact identifiers published for Z-2281-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Covidien EndoStitch" Polysorb Single Use Loading Unit; Product numbers: 170053 and 170071; | Check source notice | 10884521126732 |
| Not specified | Covidien EndoStitch" Polysorb Single Use Loading Unit; Product numbers: 170053 and 170071; | Check source notice | 10884521126787 |
FDA code information
1. Product Number: 170053; UDI-DI: 10884521126732; UPN: 20884521126739; Lot number: J5L2332Y; 2. Product Number: 170071; UDI-DI: 10884521126787; UPN: 20884521126784; Lot number: J5H2924Y;
Why was this product recalled?
The presence of holes in the pouch may compromise the package s ability to maintain the sterile barrier and protect the product from environmental exposure.
What actions did Covidien LP take?
On April 23, 2026 URGENT: VOLUNTARY MEDICAL DEVICE RECALL letters were sent to customers. Customer actions: - Immediately identify and quarantine all unused impacted product listed above. - See Appendix A for guidance on identifying affected lots. - This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Return all unused impacted product listed above to Medtronic. Medtronic Actions: - Medtronic representatives will assist customers with the return of affected product, upon request. - Medtronic representatives will work with customers to ensure replacement products are available to avoid interruptions in patient care.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2281-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology