Brand Name: Ultra IABP Catheter Kit Product Name: Ultra IABP Catheter Kit Recall
Brand Name: Ultra IABP Catheter Kit Product Name: Ultra IABP Catheter Kit Model/Catalog Number: IMU7F-40, IMU7F-35, IMU7F-30, IMU7F-25, and IMU7F-20 Software Version: Not applicable Product Description: The Insightra Ultra IABP Catheter Kit is a 7 French catheter available with either a 40, 35, 30, 25, or 20 cc balloon. The catheter has two lumens and is designed to provide counterpulsation cardiac assist therapy. The outer lumen is a channel for helium used to inflate and deflate the balloon. The inner lumen is used for a guidewire and blood pressure measurement. It is packaged as a convenience kit with accessories (one-way valve, 50 cc syringe, guide wires, introducer needle, dilator, pressure tubing, Datascope & Arrow adaptors). The balloon inflation and deflation cycle is synchronized (using a commercially available control console) with the ECG or arterial pressure to provide counterpulsation synchronized with the heartbeat. It is intended to increase coronary perfusion, decrease the workload of the left ventricle and allow healing of the myocardium Component: No
FDA recall Class II
Terminated
FDA lists this Class II recall for Brand Name: Ultra IABP Catheter Kit Product Name: Ultra IABP Catheter Kit. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2298-2025
- Recall classification
- Class II
- Enforcement status
- Terminated
- Initiation date
- Jul 19, 2025
- FDA report date
- Aug 20, 2025
- Quantity in commerce
- 24227 units
- Distribution
- See key details
- FDA cause
- No Marketing Application
What is being recalled?
Brand Name: Ultra IABP Catheter Kit Product Name: Ultra IABP Catheter Kit Model/Catalog Number: IMU7F-40, IMU7F-35, IMU7F-30, IMU7F-25, and IMU7F-20 Software Version: Not applicable Product Description: The Insightra Ultra IABP Catheter Kit is a 7 French catheter available with either a 40, 35, 30, 25, or 20 cc balloon. The catheter has two lumens and is designed to provide counterpulsation cardiac assist therapy. The outer lumen is a channel for helium used to inflate and deflate the balloon. The inner lumen is used for a guidewire and blood pressure measurement. It is packaged as a convenience kit with accessories (one-way valve, 50 cc syringe, guide wires, introducer needle, dilator, pressure tubing, Datascope & Arrow adaptors). The balloon inflation and deflation cycle is synchronized (using a commercially available control console) with the ECG or arterial pressure to provide counterpulsation synchronized with the heartbeat. It is intended to increase coronary perfusion, decrease the workload of the left ventricle and allow healing of the myocardium Component: No
Affected model & UDI checker
Search the exact identifiers published for Z-2298-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| IMU7F-40 | Brand Name: Ultra IABP Catheter Kit Product Name: Ultra IABP Catheter Kit | Check source notice | Not specified |
| IMU7F-35 | Brand Name: Ultra IABP Catheter Kit Product Name: Ultra IABP Catheter Kit | Check source notice | Not specified |
| IMU7F-30 | Brand Name: Ultra IABP Catheter Kit Product Name: Ultra IABP Catheter Kit | Check source notice | Not specified |
| IMU7F-25 | Brand Name: Ultra IABP Catheter Kit Product Name: Ultra IABP Catheter Kit | Check source notice | Not specified |
FDA code information
Lot code: All lots / UDI: IMU7F-40: 00850012832546; IMU7F-35: 00850012832515; IMU7F-30: 00850012832485; IMU7F-25: None; IMU7F-20: None
Why was this product recalled?
Device contains indications for use and device compatibility claims that have not be reviewed and approved for safety and effectiveness by the FDA. Instructions for use are not consistent with product training.
What actions did Insightra Medical Inc take?
On June 19, 2025, Insightra Medical issued a recall notification to affected consignees via postal mail and email. On August 4, 2025, Insightra Medical issued a "Urgent Medical Device Recall" notification update via E-Mail to provide additional information concerning the labeling. Insightra ask consignees to take the following actions: 1. Inform consignees to discard any previously printed versions of the IFU. 2.Review your firm s procedures and processes to ensure that they are up to date with the updated IFU. 3.If your firm has further distributed the devices to other facilities, please provide that facility with a copy of the notification. 4.Ensure that all clinicians and users are informed of the IFU updates. 5.Access the latest IFU via the QR code located inside the product box lid or via web address: https://tn-mi.com/wp-content/uploads/2023/06/D02-0043-Ultra-IAB-IFU-2019.pdf 6.Retain the letter with your product documentation. 7.Please email the acknowledgement form to us at [email protected]. 8. Report any device concerns to Insightra. Please note that adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Recall termination recorded
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2298-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026