Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2301-2026

BD Pyxis Mini Main REF: 349 Medication cabinet Recall

BD Pyxis Mini Main REF: 349 Medication cabinet

FDA recall Class II

Ongoing

FDA lists this Class II recall for BD Pyxis Mini Main REF: 349 Medication cabinet. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2301-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 30, 2026
FDA report date
Jun 10, 2026
Quantity in commerce
709 units
Distribution
See key details
FDA cause
Component design/selection

What is being recalled?

BD Pyxis Mini Main REF: 349 Medication cabinet

Affected model & UDI checker

Search the exact identifiers published for Z-2301-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
349BD Pyxis Mini Main REF: 349 Medication cabinetCheck source noticeNot specified
Not specifiedBD Pyxis Mini Main REF: 349 Medication cabinetCheck source notice10885403512582
FDA code information

All Serial Numbers/UDI:10885403512582

View the original FDA recall notice

Why was this product recalled?

Fingerprint scanner may overheat to a temperature to cause 1st degree burn.

What actions did CareFusion 303, Inc. take?

On April 30, 2026, CareFusion (subsidiary of Becton, Dickinson and Company) issued a Urgent Medical Device Recall Notification, via mail or Email. CareFusion ask consignees to take the following actions: 1. Dispensing cabinets can continue to be used as intended. 2. If you suspect that the Bio-ID Scanner is hot, please discontinue the use of Scanner and file a complaint. 3. Review and update facility policies and procedures related to situations where the Bio-ID Scanner is unavailable to mitigate delays in care. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology