Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2306-2026

MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. Electrocardiograph Recall

MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. Electrocardiograph

FDA recall Class II

Ongoing

FDA lists this Class II recall for MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. Electrocardiograph. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2306-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 21, 2026
FDA report date
Jun 10, 2026
Quantity in commerce
16
Distribution
See key details
FDA cause
Software Manufacturing/Software Deployment

What is being recalled?

MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. Electrocardiograph

Affected model & UDI checker

Search the exact identifiers published for Z-2306-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. ElectrocardiographCheck source notice00195278276070
Not specifiedMAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. ElectrocardiographCheck source notice00195278276124
FDA code information

Software version 3.00 855001-001, UDI-DI 00195278276070 8855002-001, UDI-DI 00195278276124

View the original FDA recall notice

Why was this product recalled?

Software in development inadvertently made available for customer use. Use of this software version can result in clinical information being associated with the incorrect patient. This could impact clinical decision-making.

What actions did GE Medical Systems China Co., Ltd. take?

Firm notified consignees beginning April 21, 2026 via letters titled "URGENT MEDICAL DEVICE CORRECTION". Customers were given instructions on how to check their software version. If a customer's software version begins with 3.00, the customer should not use the system until a GE HealthCare representative has updated the device with the latest licensed software version.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.