Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2313-2026

Plum Duo Precision IV Pump, 40002-0403 Recall

Plum Duo Precision IV Pump, 40002-0403

FDA recall Class II

Ongoing

FDA lists this Class II recall for Plum Duo Precision IV Pump, 40002-0403. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2313-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 29, 2026
FDA report date
Jun 17, 2026
Quantity in commerce
13,613
Distribution
See key details
FDA cause
Software Design Change

What is being recalled?

Plum Duo Precision IV Pump, 40002-0403

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

UDI-DI M335400031 Plum Duo Precision IV Pump v1.2.3

View the original FDA recall notice

Why was this product recalled?

Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Icu Medical, Inc. take?

On April 29, 2026, the firm notified consignees via letters titled "URGEND MEDICAL DEVICE CORRECTION." Customers were instructed to identify all affected pumps in their possession and ensure all users or potential users of the pumps are immediately made aware of the notification. During bolus programming, clinicians should check whether the soft limit override icon is displayed next to the bolus dose entry field on the programming screen. If the pump displays the icon, the clinician should review the program and make any necessary adjustments before confirming the program. The firm is developing a software update to correct the issue and will notify customers when it is available.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology