TDC VELOCE", 25G Two Dimensional Cutter - 20k CPM; Article Numbers: 9268.VIT25, 9311.25G01, 9311.25G02, 9311.25K01, 9311.25K02, 9311.25K03, 9625.G0201; Recall
TDC VELOCE", 25G Two Dimensional Cutter - 20k CPM; Article Numbers: 9268.VIT25, 9311.25G01, 9311.25G02, 9311.25K01, 9311.25K02, 9311.25K03, 9625.G0201;
FDA recall Class II
Ongoing
FDA lists this Class II recall for TDC VELOCE", 25G Two Dimensional Cutter - 20k CPM; Article Numbers: 9268.VIT25, 9311.25G01, 9311.25G02, 9311.25K01, 9311.25K02, 9311.25K03, 9625.G0201;. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2323-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 30, 2026
- FDA report date
- Jun 17, 2026
- Quantity in commerce
- 2068 boxes totaling 12408 units
- Distribution
- See key details
- FDA cause
- Process design
What is being recalled?
TDC VELOCE", 25G Two Dimensional Cutter - 20k CPM; Article Numbers: 9268.VIT25, 9311.25G01, 9311.25G02, 9311.25K01, 9311.25K02, 9311.25K03, 9625.G0201;
Affected model & UDI checker
Search the exact identifiers published for Z-2323-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | TDC VELOCE", 25G Two Dimensional Cutter - 20k CPM; Article Numbers: 9268.VIT25, 9311.25G01, 9311.25G02, 9311.25K01, 9311.25K02, 9311.25K03, 9625.G0201; | Check source notice | 08717872031873 |
| Not specified | TDC VELOCE", 25G Two Dimensional Cutter - 20k CPM; Article Numbers: 9268.VIT25, 9311.25G01, 9311.25G02, 9311.25K01, 9311.25K02, 9311.25K03, 9625.G0201; | Check source notice | 08717872035116 |
| Not specified | TDC VELOCE", 25G Two Dimensional Cutter - 20k CPM; Article Numbers: 9268.VIT25, 9311.25G01, 9311.25G02, 9311.25K01, 9311.25K02, 9311.25K03, 9625.G0201; | Check source notice | 08717872035345 |
FDA code information
1. Article Number: 9268.VIT25; UDI-DI: 08717872031873; 2. Article Number: 9311.25G01; UDI-DI: 08717872035116; 3. Article Number: 9311.25G02; UDI-DI: 08717872035345; 4. Article Number: 9311.25K01; UDI-DI: N/A; 5. Article Number: 9311.25K02; UDI-DI: N/A; 6. Article Number: 9311.25K03; UDI-DI: N/A; 7. Article Number: 9625.G0201; UDI-DI: N/A; Lot Numbers: 101118001, 101799001, 101800001, 102675001, 102676001, 103507001, 103508001, 103509001, 104360001, 50489-*-*-1, 54992-*-*-1, 54993-*-*-1, 54994-*-*-1, 56251-*-*-1, 56254-*-*-1, 56255-*-*-1, 56256-*-*-1, 56257-*-*-1, 57683-*-*-1, 58588-*-*-1, 58589-*-*-1, 60300-*-*-1, 65315-*-*-1, 65316-*-*-1, 65317-*-*-1, 65318-*-*-1, 65319-*-*-1, 65320-*-*-1, 65321-*-*-1, 65322-*-*-1, 66818-*-*-1, 70551-*-*-1, 70552-*-*-1, 70565-*-*-1, 70566-*-*-1, 70567-*-*-1, 71803-*-*-1, 71804-*-*-1, 71805-*-*-1, 71806-*-*-1, 75717-*-*-1, 76191-*-*-1, 76516-*-*-1, 76806-*-*-1, 76807-*-*-1, 76808-*-*-1, 76852-*-*-1, 76853-*-*-1, 76854-*-*-1, 76855-*-*-1, 77374-*-*-1, 77375-*-*-1, 77501-*-*-1, 77502-*-*-1, 77503-*-*-1, 77504-*-*-1, 77505-*-*-1, 78097-*-*-1, 78098-*-*-1, 78110-*-*-1, 78111-*-*-1, 79425-*-*-1, 79426-*-*-1, 81292-*-*-1, 81293-*-*-1, 81681-*-*-1, 82172-*-*-1, 82173-*-*-1, 82174-*-*-1, 82326-*-*-1, 82702-*-*-1, 82703-*-*-1, 82704-*-*-1, 82778-*-*-1, 83543-*-*-1, 83544-*-*-1, 84535-*-*-1, 84536-*-*-1, 84537-*-*-1, 85411001, 85412001, 86024001, 86025001, 86026001, 86806001, 86807001, 87173001, 87174001, 87175001, 87390001, 87950001, 88204001, 88205001, 88552001, 88616001, 89535001, 90377001, 90378001, 90379001, 91280001, 91281001, 92631001, 92633001, 95437001, 99667001;
Why was this product recalled?
An increase in incidence in complaints of a loose outer knife and reports of the outer knife becoming loose during use.
What actions did D.O.R.C. Dutch Opthalmic Research Center Intl B.V. take?
On April 30, 2026 Urgent Medical Device Recall letters were sent to customers. Actions to be taken: Review this Product information as soon as possible with relevant members of your staff. 1. Pass this FSN as soon as possible on to all those who need to be aware of it, including other locations, organizations and/or customers who might have received the impacted Product(s) through you. 2. If you have the affected Products and the affected lots (see attachment below) in your inventory, remove and return them to DORC following the instructions in the attachment. 3. Follow the instructions on how to complete and submit the Field Action Acknowledgement Form that is provided with this FSN.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2323-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology