FDA RECALL RECORD · Z-2328-2025

Hamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive pressure ventilatory support to adults and pediatrics and optionally infants and neonates. Recall

Hamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive pressure ventilatory support to adults and pediatrics and optionally infants and neonates.

FDA recall Class I

Ongoing

FDA lists this Class I recall for Hamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive pressure ventilatory support to adults and pediatrics and optionally infants and neonates.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2328-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Jun 3, 2025
FDA report date
Aug 27, 2025
Quantity in commerce
1040 systems
Distribution
See key details
FDA cause
Equipment maintenance

What is being recalled?

Hamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive pressure ventilatory support to adults and pediatrics and optionally infants and neonates.

Affected model & UDI checker

Search the exact identifiers published for Z-2328-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
160021Hamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive pressure ventilatory support to adults and pediatrics and optionally infants and neonates.Check source noticeNot specified
Not specifiedHamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive pressure ventilatory support to adults and pediatrics and optionally infants and neonates.Check source notice07630002808590
FDA code information

Model/Catalog Number: 160021 ESM board Part Number: 10144626 GTIN: 07630002808590 UDI-DI: 76300028PN160021ZW Affected SN range: 13636 till 13950 and 13966 till 14753 13797 13801 13806 13855 13865 13981 13988 13989 14012 14015 14019 14034 14035 14042 14052 14054 14102 14109 14112 14113 14114 14115 14116 14119 14120 14121 14122 14123 14124 14126 14127 14128 14163 14168 14169 14170 14172 14173 14175 14177 14187 14188 14189 14195 14197

View the original FDA recall notice

Why was this product recalled?

A mechanical defect on a printed circuit board (PCB) as a result of PCB separating process. This defect might cause the ventilator to either fail at startup (no patient involvement) or during the use of the device enter ambient state, resulting in an interruption of ventilation.

What actions did Hamilton Medical AG take?

URGENT: Medical Device Correction letters were emailed to customers on 06/06/2025, 06/20/2025, and 07/07/2025, informing them that due to damage to the capacitor in the ESM-Board within the Hamilton-C6 ventilator the capacitor could, over time, lead to a loss of its function due to oxidation of similar degradation processes. Customers are instructed to: 1. Review their Hamilton-C6 ventilators and check for the serial number within the scope mentioned in the letter. 2. for affected ventilators must have the ESM-Boards replaced. The replacement will be organized by the local partner as per servicing protocol. Note - the Hamilton-C6 ventilators do not need to be removed from use/service. These devices were tested prior to release and no test will be able to detect if the ESM board will fail in the future. 3. Keep the recall notice within the respective ventilator's data records. For questions or assistance, contact Technical Support at 1-800-426-6331 option 2 or email [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology