FDA RECALL RECORD · Z-2332-2026

Brand Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS Product Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS Recall

Brand Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS Product Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS Model/Catalog Number: 1023770 Software Version: N/A Product Description: The SporeTraq Mail-In Spore Testing Service includes 12 bacterial Spore Strips provided by Mesa Labs. The Spore Strip (single species G. stearothermophilus ATCC#7953 or dual species G. stearothermophilus ATCC #7953 and B. atrophaeus ATCC #9372) is intended for use in testing the efficacy of sterilizer systems used in healthcare facilities. The Spore Strips are compatible with Steam sterilization, Chemiclave¿ sterilization, Ethylene Oxide sterilization, or Dry Heat sterilization processes. The test strip is placed in the customer s sterilizer during their sterilization cycle and upon completion the sterilized test strip along with the control strip is returned to Mesa Labs for lab testing in the return envelope. Spore test results are digitally stored in HealthFirst s OnTraq smart automation tracking and reporting system for easy and secure access to a paperless archive from any device with internet access. Component: N/A

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS Product Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2332-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 13, 2026
FDA report date
Jun 17, 2026
Quantity in commerce
0
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Brand Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS Product Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS Model/Catalog Number: 1023770 Software Version: N/A Product Description: The SporeTraq Mail-In Spore Testing Service includes 12 bacterial Spore Strips provided by Mesa Labs. The Spore Strip (single species G. stearothermophilus ATCC#7953 or dual species G. stearothermophilus ATCC #7953 and B. atrophaeus ATCC #9372) is intended for use in testing the efficacy of sterilizer systems used in healthcare facilities. The Spore Strips are compatible with Steam sterilization, Chemiclave sterilization, Ethylene Oxide sterilization, or Dry Heat sterilization processes. The test strip is placed in the customer s sterilizer during their sterilization cycle and upon completion the sterilized test strip along with the control strip is returned to Mesa Labs for lab testing in the return envelope. Spore test results are digitally stored in HealthFirst s OnTraq smart automation tracking and reporting system for easy and secure access to a paperless archive from any device with internet access. Component: N/A

Affected model & UDI checker

Search the exact identifiers published for Z-2332-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
1023770Brand Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS Product Name: SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTSCheck source noticeNot specified
FDA code information

Lot Code: Please see attached spreadsheet listing products that were not incubated per the following requirements : a. Spore strips used in steam sterilizers are incubated for 24 hours b. Spore strips used in chemiclave sterilizers are incubated for 72 hours c. Spore strips used in dry heat or ethylene oxide sterilizers are incubated for 7 days As noted in the previous recall scope question, HealthFirst's Field Correction scope will be expanded to include notifications to all customers in possession of 1023780 SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 52 TESTS,1023770 SPORETRAQ MAIL-IN SPORE TESTING SERVICE, 12 TESTS or other products serviced by Mesa Labs.

View the original FDA recall notice

Why was this product recalled?

Due to incorrect incubation process being performed on product.

What actions did HF Acquisition Co LLC take?

On 04/13/2026, the firm emailed an "URGENT: Medical Device Field Correction" Letter to customers informing them that the firm has identified a laboratory testing service error associated with 1023780 SporeTraq" Mail-In Spore Testing Service, 52 Tests affecting customers using Dry Heat, Chemiclave and Ethylene Oxide autoclave types. Specifically, the incubation criteria applied to the test samples may not have complied with the above requirement for the respective autoclave type. Failure to adhere to appropriate incubation criteria impacts the validity of sterilization process followed at customer locations. Customers are instructed to: 1. Log into your OnTraq" account at ontraq.healthfirst.com. 2. A banner will appear stating ACTION REQUIRED: Review to confirm or update your autoclave types. Go to the Confirmations Page. prompting you to verify the autoclave type. 3. Verify that the autoclave type is accurate by clicking the "Go to the Confirmations Page" hyperlink or navigating to "Confirm Autoclave Type" tab - If all information is accurate as it is, click on the CONFIRM button. - If not, edit by clicking the pencil icon next to the relevant data field, enter the updated information and click on the SAVE button. - If the autoclave is no longer in use, click the DEACTIVATE button and confirm the action in the pop-up window. 4. If the above actions were completed prior to receipt of this notification via email, mail or telephone, no further action is required. For questions - contact Customer Service at 855-476-1342 or email us at [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.