Estrone RIA Catalog Number DSL8700 UDI Code: 15099590211615 Estrone RIA is an in vitro diagnostic manual medical device used by healthcare professionals for the quantitative measurement of estrone in human serum and plasma. Recall
Estrone RIA Catalog Number DSL8700 UDI Code: 15099590211615 Estrone RIA is an in vitro diagnostic manual medical device used by healthcare professionals for the quantitative measurement of estrone in human serum and plasma.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Estrone RIA Catalog Number DSL8700 UDI Code: 15099590211615 Estrone RIA is an in vitro diagnostic manual medical device used by healthcare professionals for the quantitative measurement of estrone in human serum and plasma.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2336-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jun 12, 2025
- FDA report date
- Aug 27, 2025
- Quantity in commerce
- 279 units
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Estrone RIA Catalog Number DSL8700 UDI Code: 15099590211615 Estrone RIA is an in vitro diagnostic manual medical device used by healthcare professionals for the quantitative measurement of estrone in human serum and plasma.
Affected model & UDI checker
Search the exact identifiers published for Z-2336-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| DSL8700 | Estrone RIA Catalog Number DSL8700 UDI Code: 15099590211615 Estrone RIA is an in vitro diagnostic manual medical device used by healthcare professionals for the quantitative measurement of estrone in human serum and plasma. | Check source notice | Not specified |
| Not specified | Estrone RIA Catalog Number DSL8700 UDI Code: 15099590211615 Estrone RIA is an in vitro diagnostic manual medical device used by healthcare professionals for the quantitative measurement of estrone in human serum and plasma. | Check source notice | 15099590211615 |
FDA code information
Catalog Number DSL8700 UDI Code: 15099590211615 Serial/Lot Numbers: 241014D 241111C 241209C 241209D 250106C 250106D 250203C
Why was this product recalled?
Estrone RIA kit lots may rarely give falsely high patient results which might cause Temporary injury of Remote probability or rarely give falsely low patient results which might cause Permanent injury of Remote probability. Only in some (very rare) cases may the issue occur.
What actions did Immunotech A.S. take?
On 06/12/2025, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter to customers informing them that Beckman Coulter determined that the Estrone RIA kit might be affected by an isolated coated tubes inhomogeneity issue causing falsely elevated results or falsely decreased results. Customers are instructed to: Beckman Coulter recommends sharing the content of the Recall Letter with their laboratory and/or medical director regarding the need to review previous patient test results. Note- All affected product have expired (expiry dates 12/31/2024 thru 04/22/2025) For questions, contact Customer Support Center: Website: http://www.beckmancoulter.com Email: [email protected]
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2336-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026