Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2413-2026

Orca Single Use Air/Water and Suction Valves, Material Number (UPN) SUV-617-50; The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI (gastrointestinal) endoscopic procedure. Recall

Orca Single Use Air/Water and Suction Valves, Material Number (UPN) SUV-617-50; The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI (gastrointestinal) endoscopic procedure.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Orca Single Use Air/Water and Suction Valves, Material Number (UPN) SUV-617-50; The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI (gastrointestinal) endoscopic procedure.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2413-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 12, 2026
FDA report date
Jun 17, 2026
Quantity in commerce
9750 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Orca Single Use Air/Water and Suction Valves, Material Number (UPN) SUV-617-50; The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI (gastrointestinal) endoscopic procedure.

Affected model & UDI checker

Search the exact identifiers published for Z-2413-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedOrca Single Use Air/Water and Suction Valves, Material Number (UPN) SUV-617-50; The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI (gastrointestinal) endoscopic procedure.Check source notice00840253111012
FDA code information

GTIN 00840253111012, Lot 38400303, Exp. January 20, 2029

View the original FDA recall notice

Why was this product recalled?

Boston Scientific is initiating a medical device removal of Orca Sterile, Single Use Air/Water and Suction Valves due to a higher than anticipated occurrence of the suction button sticking.

What actions did BOSTON SCIENTIFIC CORPORATION take?

Boston Scientific issued an Urgent Medical Device Removal - Immediate Action Required notice to its consignees on 5/12/2026 via FedEx. The notice explained the issue with the device, potential risk to the patient, and requested the following: Actions " Immediately stop further use or distribution and segregate affected product. " Complete and return the enclosed Reply Verification Tracking Form per the included instructions. " Return product to Boston Scientific in accordance with the enclosed instructions. " Share this communication with any healthcare professionals in your hospital that use the product and with any other organization to which these systems may have been transferred. For additional assistance or more information regarding this communication, please contact your local Boston Scientific representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.