Orca Single Use Air/Water and Suction Valves, Material Number (UPN) SUV-617-50; The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI (gastrointestinal) endoscopic procedure. Recall
Orca Single Use Air/Water and Suction Valves, Material Number (UPN) SUV-617-50; The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI (gastrointestinal) endoscopic procedure.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Orca Single Use Air/Water and Suction Valves, Material Number (UPN) SUV-617-50; The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI (gastrointestinal) endoscopic procedure.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2413-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- May 12, 2026
- FDA report date
- Jun 17, 2026
- Quantity in commerce
- 9750 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Orca Single Use Air/Water and Suction Valves, Material Number (UPN) SUV-617-50; The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI (gastrointestinal) endoscopic procedure.
Affected model & UDI checker
Search the exact identifiers published for Z-2413-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Orca Single Use Air/Water and Suction Valves, Material Number (UPN) SUV-617-50; The Orca Air/Water Valve is intended to be used to control the air/water function on an endoscope during a GI (gastrointestinal) endoscopic procedure. | Check source notice | 00840253111012 |
FDA code information
GTIN 00840253111012, Lot 38400303, Exp. January 20, 2029
Why was this product recalled?
Boston Scientific is initiating a medical device removal of Orca Sterile, Single Use Air/Water and Suction Valves due to a higher than anticipated occurrence of the suction button sticking.
What actions did BOSTON SCIENTIFIC CORPORATION take?
Boston Scientific issued an Urgent Medical Device Removal - Immediate Action Required notice to its consignees on 5/12/2026 via FedEx. The notice explained the issue with the device, potential risk to the patient, and requested the following: Actions " Immediately stop further use or distribution and segregate affected product. " Complete and return the enclosed Reply Verification Tracking Form per the included instructions. " Return product to Boston Scientific in accordance with the enclosed instructions. " Share this communication with any healthcare professionals in your hospital that use the product and with any other organization to which these systems may have been transferred. For additional assistance or more information regarding this communication, please contact your local Boston Scientific representative.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2413-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology