Spectrum IQ Infusion Pump, Product Code 3570009 Recall
Spectrum IQ Infusion Pump, Product Code 3570009
FDA recall Class II
Ongoing
FDA lists this Class II recall for Spectrum IQ Infusion Pump, Product Code 3570009. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2415-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 14, 2025
- FDA report date
- Sep 3, 2025
- Quantity in commerce
- 6 units
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Spectrum IQ Infusion Pump, Product Code 3570009
Affected model & UDI checker
Search the exact identifiers published for Z-2415-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Spectrum IQ Infusion Pump, Product Code 3570009 | Check source notice | 00085412610900 |
FDA code information
UDI-DI: 00085412610900; Serial Numbers: 3815255 3815261 3815265 3815266 3815267 3815271
Why was this product recalled?
Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Baxter Healthcare Corporation take?
Customers were contacted via telephone beginning 7/14/25. The included information about the product, the issue and the action to be taken by the customers. Baxter Healthcare is requesting the return of the impacted product to perform all tests required and confirm that the product perform as expected. If you have any questions, phone (Toll Free): 800-422-9837 or 847-948-4770; hours of operation: Monday through Friday from 8:00 am to 5:00 pm CST
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2415-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.