Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2418-2025

Brand Name: THOR Product Name: NovoTHOR Gen 3.0 Reg PRX Recall

Brand Name: THOR Product Name: NovoTHOR Gen 3.0 Reg PRX Model/Catalog Number: S2189 Software Version: N/A Product Description: The NovoTHOR and NovoTHOR XL are intended to emit energy in the visible and IR spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary increase in local blood circulation, temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, muscle spasm, and temporary relaxation of muscles. Component: N/A

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: THOR Product Name: NovoTHOR Gen 3.0 Reg PRX. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2418-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 24, 2025
FDA report date
Sep 3, 2025
Quantity in commerce
1 unit
Distribution
See key details
FDA cause
Process control

What is being recalled?

Brand Name: THOR Product Name: NovoTHOR Gen 3.0 Reg PRX Model/Catalog Number: S2189 Software Version: N/A Product Description: The NovoTHOR and NovoTHOR XL are intended to emit energy in the visible and IR spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary increase in local blood circulation, temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, muscle spasm, and temporary relaxation of muscles. Component: N/A

Affected model & UDI checker

Search the exact identifiers published for Z-2418-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
S2189Brand Name: THOR Product Name: NovoTHOR Gen 3.0 Reg PRXCheck source noticeNot specified
Not specifiedBrand Name: THOR Product Name: NovoTHOR Gen 3.0 Reg PRXCheck source notice05060494130480
FDA code information

Model/Catalog Number: S2189; UDI-DI: 05060494130480; Only one device of this type is in scope. Serial number: 4068;

View the original FDA recall notice

Why was this product recalled?

The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did THOR Photomedicine Ltd take?

On July 24, 2025, IMPORTANT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions by THOR: A service agent representing THOR will contact you within the next 10 weeks to schedule the fitting of the four high tensile strength steel bolts to the inside of the chassis. The process is likely to take around 3 hours and, of course, there is no fee. Required Actions for NovoTHOR Owners: 1. Please reply to this email within four days to acknowledge your receipt and understanding of this notification. 2. If you no longer own the NovoTHOR, please forward this notice to the new owner if the device was sold and reply to this email with the contact details of the new owner. Other Actions for Owners: Should you continue to have users/patients use the NovoTHOR, then be aware it may potentially cause harm should the failure occur whilst they are lifting or lowering the canopy. " Ensure you warn all users of the potential risk " Ensure someone is present to lift and lower the canopy " Ensure the canopy is lifted and lowered gently and slowly " Ensure someone is present in the room or immediately outside the room during use. THOR Contact Information: Call +44 (0)1494 797 100, Monday through Friday, 5:00 AM to 2:00 PM, Eastern Time or via thorlaser.com/contact.htm.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology