Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2422-2025

Brand Name: Atlantis IO FLO Product Name: Atlantis IO FLO-S Kit Recall

Brand Name: Atlantis IO FLO Product Name: Atlantis IO FLO-S Kit Model/Catalog Number: A04S / 6802034 Product Description: The Atlantis intraoral feature locating object for suprastructures (IO FLO-S) is intended to support the prosthetic procedure with the aim of restoring chewing function in partial or complete edentulous patients, by enabling digital impression taking on implant or abutment level for all positions in the mouth of the patient or on the master model. Component: not a component

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Atlantis IO FLO Product Name: Atlantis IO FLO-S Kit. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2422-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 24, 2025
FDA report date
Sep 3, 2025
Quantity in commerce
780 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Brand Name: Atlantis IO FLO Product Name: Atlantis IO FLO-S Kit Model/Catalog Number: A04S / 6802034 Product Description: The Atlantis intraoral feature locating object for suprastructures (IO FLO-S) is intended to support the prosthetic procedure with the aim of restoring chewing function in partial or complete edentulous patients, by enabling digital impression taking on implant or abutment level for all positions in the mouth of the patient or on the master model. Component: not a component

Affected model & UDI checker

Search the exact identifiers published for Z-2422-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
A04SBrand Name: Atlantis IO FLO Product Name: Atlantis IO FLO-S KitCheck source noticeNot specified
Not specifiedBrand Name: Atlantis IO FLO Product Name: Atlantis IO FLO-S KitCheck source notice07392532249440
FDA code information

Catalog / Ref #: 68020034; UDI: 07392532249440; Lot #s: 10515863, 10512084, 10505543, 10556741;

View the original FDA recall notice

Why was this product recalled?

The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to full engagement which is approximately 5 rotations. This equates to approximately 1mm difference in depth of engagement.

What actions did Dentsply IH, Inc. take?

On July 24, 2025, URGENT MEDICAL DEVICE RECALL letters were sent to customers. Action to be taken by the Customer: 1. Stop use of the affected lots immediately. 2. If the IO FLO-S has already been used for scanning, please immediately contact the dental laboratory that manufactured the dental restoration and direct the laboratory to contact Dentsply Sirona Atlantis Customer Service, immediately (see contact data below) to verify if the scans are correct. 3. If the IO FLO-S has already been used and a structure has been manufactured and placed into the patient, please immediately contact the patient and verify the correct engagement of the suprastructure, correct fit and torque of the prosthetic screws. Please contact your local Dentsply Sirona Atlantis Customer Service, immediately (see contact data below). 4. Return any affected lots for refund or replacement to the address below by including the completed and signed Answer Letter (see page 3). If unable to identify the affected lots, please return all products of the affected types. Dentsply IH Inc Attn: QA Department 590 Lincoln St Waltham, MA 02451 5. Please return the completed and signed Answer Letter (page 3 of this document) to the local Dentsply Sirona country organization identified above within one week after receipt of this Urgent Medical Device Recall Notice. Timely completion will allow us to process your replacement product more quickly. Transmission of this Urgent Medical Device Recall Notice Please provide this Notice to everyone that needs to be aware of this issue inside or outside of your organization. Specifically, please provide this Notice to any organization where the potentially affected product(s) have been transferred or used within the workflow to create a patient-specific restoration, or where this issue may have an impact. Please continue to review and abide by this Notice for an adequate timeframe so all affected products are identified, and appropriate measure are taken Please repor

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology