Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2446-2025

Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App Recall

Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App Model/Catalog Number: SW12299 Software Version: version 2.8.0 and earlier Product Description: Dexcom G7 CGM Android App SW12299 Component: Dexcom G7 Continuous Glucose Monitoring System

FDA recall Class I

Ongoing

FDA lists this Class I recall for Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2446-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Jul 24, 2025
FDA report date
Sep 10, 2025
Quantity in commerce
532821
Distribution
See key details
FDA cause
Software design

What is being recalled?

Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App Model/Catalog Number: SW12299 Software Version: version 2.8.0 and earlier Product Description: Dexcom G7 CGM Android App SW12299 Component: Dexcom G7 Continuous Glucose Monitoring System

Affected model & UDI checker

Search the exact identifiers published for Z-2446-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
SW12299Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android AppCheck source noticeNot specified
FDA code information

Lot Code: UDI: SW12299 - 00386270001870

View the original FDA recall notice

Why was this product recalled?

The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the App. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a "transmitterFailed" error message to the App when the CGM experiences a hardware or firmware failure. Instead, the App ends the CGM sensor session, stops reporting glucose values and displays the Start Sensor screen or No active sensor message without alerting the user, which can lead to missed detection of a hyperglycemic or hypoglycemic event and a delay in treatment.

What actions did Dexcom, Inc. take?

On 07/24/2025, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION NOTIFICATION" via in-app messaging informing users that a software bug was identified with users using the following app version: -Dexcom ONE+ iOS/Android App version 1.4 and all earlier versions, -Dexcom G7 iOS/G7 Watch iOS/Android App versions 2.8 and earlier The software bug can cause an issue where, when a transmitter error occurs, the app terminates the sensor session and prompts the user to "start a new sensor" without providing a "sensor failed" alert. Customer are instructed that an upgrade from app version version the affected software versions is mandatory and users will not be able to to use the affected app versions after August 20, 2025. To update their app now to continue to use the app: 1. Tap Update App to go to the app store. 2. Install the latest Dexcom ONE+ app version or Dexcom G7 app version 3. Open the Dexcom ONE+ app or the G7 app If assistance is needed, contact Technical Support U.S.: 1-844-478-1600 Global: Find local Technical Support contact information at Dexcom.com

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology