ALYON Surgical Lighting System Recall

ALYON Surgical Lighting System

FDA recall Class II

Ongoing

FDA lists this Class II recall for ALYON Surgical Lighting System. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2450-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 15, 2026
FDA report date
Jun 24, 2026
Quantity in commerce
549 units
Distribution
See key details
FDA cause
Material/Component Contamination

What is being recalled?

ALYON Surgical Lighting System

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

UDIs: 00724995223182 00724995223137, 00724995223151, 00724995223175, 00724995222895, 00724995222994 00724995223014, 00724995223021, 00724995223083, 00724995223106, 00724995222758 00724995223120, 00724995223113, 00724995222840, 00724995223038, 00724995223076 00724995222987, 00724995223007, 00724995222703, 00724995222727, 00724995223090 00724995225711, 00724995225681, 00724995225698, 00724995225728, 00724995222826 00724995222802, 00724995222857, 00724995222697, 00724995222956, and 00724995223229

View the original FDA recall notice

Why was this product recalled?

The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach.

What actions did STERIS CORPORATION take?

An URGENT MEDICAL DEVICE RECALL notification letter dated 5/1/26 was sent to customers. STERIS Action All affected Customers will be contacted by a STERIS Service Technician to arrange an onsite inspection of their lighting system(s) to perform the correction as soon as possible. For the paint issue, each affected lighting system will undergo a full visual inspection for any signs of paint delamination; as needed a service kit will be installed to prevent paint chipping from occurring or if deemed necessary replaced. For the button issue, each affected lighting system will undergo inspection for any signs of potential separation from the light head and handle. Buttons failing this inspection will be replaced. User Action Users can continue to use their lighting system(s) prior to completion of the correction. Users are reminded to ensure no excess fluids remain on light head during or after cleaning, per the Operator Manuals. Should any issues be observed prior to the correction being completed, users should submit a complaint by contacting STERIS Customer Service at 1-800-548-4873. Users are also reminded to report any adverse events via FDA s MedWatch Database should an event occur. We apologize for any inconvenience this matter may cause, and as always, STERIS is dedicated to supporting our products and valued Customers. If you have questions regarding this matter, please contact your local Account Manager.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.