FDA RECALL RECORD · Z-2456-2026

TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816 Recall

TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816

FDA recall Class II

Ongoing

FDA lists this Class II recall for TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2456-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Mar 16, 2026
FDA report date
Jun 24, 2026
Quantity in commerce
5
Distribution
See key details
FDA cause
Process control

What is being recalled?

TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816

Affected model & UDI checker

Search the exact identifiers published for Z-2456-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
71760816TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816Check source noticeNot specified
Not specifiedTRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816Check source notice03596010514394
FDA code information

UDI-DI: 03596010514394. Lot/Batch: 25BM03616

View the original FDA recall notice

Why was this product recalled?

Humeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equivalent backup can be used, or a bent nail of the same size, which would require additional surgical steps to address soft tissue/bone preparation, which could lead to a surgical delay of greater than 30 minutes.

What actions did Smith & Nephew, Inc. take?

On 3/16/2026, recall notices were emailed to customers who were asked to do the following: 1) Locate and quarantine affected devices immediately. If you have further distributed the product to other organizations, please inform them at once of this Field Action and provide to them a copy of this letter. 3. Please complete the Customer Response form and email or fax it to your national Smith+Nephew agency/distributor. 4. Return quarantined product to your national Smith+Nephew agency/distributor. 5. Please maintain awareness of this notice and resulting action for an appropriate period to ensure the effectiveness of the corrective action. Firm contact information: + 1 901 396 2121, 1 800 821 5700

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology