Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2505-2025

Brand Name: MediHoney Product Name: MediHoney Paste Recall

Brand Name: MediHoney ¿ Product Name: MediHoney¿ Paste Model/Catalog Number: 31505 Product Description: MediHoney¿ Paste with Active Leptospermum Honey is a moist dressing made of 100% Active Leptospermum medical grade Honey. The dressing helps maintain a moist environment conducive to wound healing and helps aid and support autolytic debridement. The dressing can be removed without causing trauma to the wound when a primary dressing is used.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: MediHoney Product Name: MediHoney Paste. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2505-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 5, 2025
FDA report date
Sep 10, 2025
Quantity in commerce
15441 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Brand Name: MediHoney Product Name: MediHoney Paste Model/Catalog Number: 31505 Product Description: MediHoney Paste with Active Leptospermum Honey is a moist dressing made of 100% Active Leptospermum medical grade Honey. The dressing helps maintain a moist environment conducive to wound healing and helps aid and support autolytic debridement. The dressing can be removed without causing trauma to the wound when a primary dressing is used.

Affected model & UDI checker

Search the exact identifiers published for Z-2505-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
31505Brand Name: MediHoney Product Name: MediHoney PasteCheck source notice10381780486930
FDA code information

Model/Catalog Number: 31505; UDI: 10381780486930; Lot #s: 2150, 2336, 2151, 2337, 2201, 2414, 2209, 2415, 2217, 2417, 2232, 2428, 2241, 2434, 2301, 2435, 2312, 2505, 2313, 2508, 2314, 2509, 2328;

View the original FDA recall notice

Why was this product recalled?

Packaging failures were identified which could lead to a breach in the sterile barrier.

What actions did Integra LifeSciences Corp. (NeuroSciences) take?

On August 5, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions for Customers: If you have units of the impacted product remove them immediately from service and quarantine them. Forward this notice to those who utilize the product so they are aware of this recall and can identify any affected product that may remain in clinical areas. Complete the entirety of the Acknowledgement Form and send via email to [email protected] or FAX to 1-609-750-4220. Keep a copy of the form for your records. Actions for Distributors: If you have products listed in Table 1, remove the product from further distribution. Complete the entirety of the Acknowledgement Form and send via email to [email protected] or FAX to 1-609-750-4220. Keep a copy of the form for your records. Check your customer traceability records and notify them if they have any shipments of above catalog and lot numbers. Modify the acknowledgement form to create one from you to your customers. Collect completed response forms and affected product from your customers and indicate total quantities and lots in distributor reply form. For questions regarding these instructions, please contact Customer Service: Monday - Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873 email: [email protected] Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Integra is dedicated to patient safety and manufacturing excellence. We continue to make numerous quality improvements and investments in our facilities around the world. In addition, we are improving our processes in accordance with applicable regulations. We sincerely apologize for any inconvenience this voluntary recall may cause and thank you for your cooperation in this effort.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.