Product Name: STAAR ICL Calculation Software Software Version: Version 8.00 Product Description: The STAAR ICL Calculation Software is designed as an aide in the calculation of the size and diopter power (with residual refraction) for physician selection of the lens. For toric ICLs, an Implantation Orientation Diagram (IOD) is also produced to provide the physician with pictorial representation of the axis of rotation and alignment. Note: The STAAR ICL Calculation Software resides on STAAR Surgical's e-commerce website in STAAR ICL planning support (Stella), as referenced in the attached recall communication. Component: No Recall
Product Name: STAAR ICL Calculation Software Software Version: Version 8.00 Product Description: The STAAR ICL Calculation Software is designed as an aide in the calculation of the size and diopter power (with residual refraction) for physician selection of the lens. For toric ICLs, an Implantation Orientation Diagram (IOD) is also produced to provide the physician with pictorial representation of the axis of rotation and alignment. Note: The STAAR ICL Calculation Software resides on STAAR Surgical's e-commerce website in STAAR ICL planning support (Stella), as referenced in the attached recall communication. Component: No
FDA recall Class II
Ongoing
FDA lists this Class II recall for Product Name: STAAR ICL Calculation Software Software Version: Version 8.00 Product Description: The STAAR ICL Calculation Software is designed as an aide in the calculation of the size and diopter power (with residual refraction) for physician selection of the lens. For toric ICLs, an Implantation Orientation Diagram (IOD) is also produced to provide the physician with pictorial representation of the axis of rotation and alignment. Note: The STAAR ICL Calculation Software resides on STAAR Surgical's e-commerce website in STAAR ICL planning support (Stella), as referenced in the attached recall communication. Component: No. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2509-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 20, 2026
- FDA report date
- Jun 24, 2026
- Quantity in commerce
- 543
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
Product Name: STAAR ICL Calculation Software Software Version: Version 8.00 Product Description: The STAAR ICL Calculation Software is designed as an aide in the calculation of the size and diopter power (with residual refraction) for physician selection of the lens. For toric ICLs, an Implantation Orientation Diagram (IOD) is also produced to provide the physician with pictorial representation of the axis of rotation and alignment. Note: The STAAR ICL Calculation Software resides on STAAR Surgical's e-commerce website in STAAR ICL planning support (Stella), as referenced in the attached recall communication. Component: No
Affected model & UDI checker
Search the exact identifiers published for Z-2509-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Product Name: STAAR ICL Calculation Software Software Version: Version 8.00 Product Description: The STAAR ICL Calculation Software is designed as an aide in the calculation of the size and diopter power (with residual refraction) for physician selection of the lens. For toric ICLs, an Implantation Orientation Diagram (IOD) is also produced to provide the physician with pictorial representation of the axis of rotation and alignment. Note: The STAAR ICL Calculation Software resides on STAAR Surgical's e-commerce website in STAAR ICL planning support (Stella), as referenced in the attached recall communication. Component: No | Check source notice | 00840311346288 |
FDA code information
Lot Code: UDI (01)00840311346288(0812)8.00 Software Version 8.00 The software error occurs if a customer generated an IOD for a toric implantation from April 13, 2026 to April 16, 2026 for multiple toric lens reservations for the same eye using the STAAR ICL Calculation Software.
Why was this product recalled?
Due to a software coding issues for calculating the implant orientation diagram printout for the implantation for toric implantable collamer lenses.
What actions did Staar Surgical AG take?
On 04/20/2026, the firm sent via FedEx an "URGENT: MEDICAL DEVICE RECALL" Letter informing customers that STAAR has confirmed that during that period, the printing of the IOD within the Stella Patient View, Order Confirmation, or Shopping Cart, when performed locally (i.e. by doctor/clinic), was under certain circumstances producing an incorrect diagram. The incorrect diagram was produced in certain circumstances when users created multiple Toric lens line items with different axis values on the same patient eye, such as with backup or secondary lens reservations. The root cause has been identified as a software coding error in Stella 2.0 that did not assign a unique identifier to each reservation under a patient s eye. STAAR has modified the software code to assign a unique identifier, thereby preventing this issue when multiple reservations exist. Any IOD printed from Stella 2.0 after 5:30pm PDT on April 16, 2026 would not have been affected by this issue. For Customer's awareness: We are informing you so that a proper clinical decision can be made regarding patient management and follow-up, based on your assessment of the clinical outcome. Please immediately complete and return the Recall Response Form acknowledging receipt of this recall and report any adverse events associated with this event. If a Toric EVO/EVO+ device were implanted using an incorrect IOD printout, there is the potential for the following adverse events: " Refractive surprise " Blurred vision " Glares/Halos " Other visual disturbances Repositioning or removal of the lens may need to be considered in these cases. For Questions - contact STAAR Sales Representative or Customer Service at [email protected] or by phone at 1-800-352-7842 Monday through Friday, 6:00am to 5:00pm PDT.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA classification recorded
FDA notice posted
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2509-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology