FDA lists this Class II recall for iLet Bionic Pancreas, REF: BB1001. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2513-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 23, 2025
FDA report date
Jun 24, 2026
Quantity in commerce
4580
Distribution
See key details
FDA cause
Software design

What is being recalled?

iLet Bionic Pancreas, REF: BB1001

Affected model & UDI checker

Search the exact identifiers published for Z-2513-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
BB1001iLet Bionic Pancreas, REF: BB1001Check source noticeNot specified
FDA code information

iLet ACE Pump firmware BOM000072 Rev C (version 3.3.6) and BOM000072 Rev B (version 3.3.4)

View the original FDA recall notice

Why was this product recalled?

The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could occur when the user successfully slides the unlock slider on the Lock Screen, no screen transition occurs, and the user remains on the Lock Screen. Because the user cannot exit the Lock Screen, the user is unable to perform a cartridge change or meal announcement. If this occurs and therapy is in progress, therapy will continue normally until the insulin cartridge is depleted. The user will be unable to perform a cartridge change or a meal announcement and will be stuck on the screen until the battery fully drains. This can lead to hyperglycemia.

What actions did Beta Bionics, Inc. take?

On 3/12/2026, recall notices were mailed and emailed to customers who were informed of the following: On September 23, 2025, Beta Bionics, Inc. issued an Urgent Medical Device Correction notification to customers informing them of the lock screen software issue and software upgrade. These problems are resolved in newer host software versions (3.3.7 and up). Please check your software version and update your software as soon as possible. Update your software 1. Download or update your iLet app on your phone. Need help! Read the iLet App User Guide. 2. Log into your iLet app. 3. Position your iLet next to your phone with the app open. 4. Go to app settings on the iLet app and follow the on-screen instructions to update to the latest firmware. 5. Complete the acknowledgement survey at https://sprw.io/stt-3tkjI or If you have any questions, please call the firm's Customer Care at 1-855-745-3800 option 1.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology