FDA RECALL RECORD · Z-2514-2025

NobelBiocare Tri-Lobe-compatible WP X 6.0mm Engaging Titanium Base, REF: 9000662 Recall

NobelBiocare Tri-Lobe-compatible WP X 6.0mm Engaging Titanium Base, REF: 9000662

FDA recall Class II

Ongoing

FDA lists this Class II recall for NobelBiocare Tri-Lobe-compatible WP X 6.0mm Engaging Titanium Base, REF: 9000662. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2514-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jun 25, 2025
FDA report date
Sep 10, 2025
Quantity in commerce
11
Distribution
See key details
FDA cause
Process control

What is being recalled?

NobelBiocare Tri-Lobe-compatible WP X 6.0mm Engaging Titanium Base, REF: 9000662

Affected model & UDI checker

Search the exact identifiers published for Z-2514-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
9000662NobelBiocare Tri-Lobe-compatible WP X 6.0mm Engaging Titanium Base, REF: 9000662Check source noticeNot specified
Not specifiedNobelBiocare Tri-Lobe-compatible WP X 6.0mm Engaging Titanium Base, REF: 9000662Check source notice00842092169599
FDA code information

UDI-DI: 00842092169599. Lot: 277097

View the original FDA recall notice

Why was this product recalled?

Engaging titanium base was manufactured 180 degrees around the post from where it should be located in relation to the connection geometry; if the flat is on the wrong side of the post, the crown will not fit onto the abutment in the correct orientation, resulting in restoration delay.

What actions did Preat Corp take?

On 06/25/2025, recall notices were mailed to customers who were asked to do the following: 1) Stop/cease use of the product in the scope of this recall 2) This product shall be discarded or returned to the recalling firm 2) All personnel within your company should be made aware of the issue. Any customers to which this product has been further distributed should be notified be provided a copy of this notification and informed to notify their customers. 3) Complete and return the Customer Response Form via email to [email protected] If you have any questions or concerns, please call/email the firm at: (800) 232-7732, [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.