TOPS Inserter, part of the TOPS System Instrument Set used for implantation of the TOPS System. Recall
TOPS Inserter, part of the TOPS System Instrument Set used for implantation of the TOPS System. Model Number: 82889.
FDA recall Class II
Ongoing
FDA lists this Class II recall for TOPS Inserter, part of the TOPS System Instrument Set used for implantation of the TOPS System.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2516-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 16, 2025
- FDA report date
- Sep 10, 2025
- Quantity in commerce
- 30 units
- Distribution
- See key details
- FDA cause
- Process change control
What is being recalled?
TOPS Inserter, part of the TOPS System Instrument Set used for implantation of the TOPS System. Model Number: 82889.
Affected model & UDI checker
Search the exact identifiers published for Z-2516-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 82889 | TOPS Inserter, part of the TOPS System Instrument Set used for implantation of the TOPS System. | Check source notice | Not specified |
| Not specified | TOPS Inserter, part of the TOPS System Instrument Set used for implantation of the TOPS System. | Check source notice | 07290115932638 |
FDA code information
Model Number: 82889. UDI-DI: 07290115932638. Lot Numbers: RS20220546-1, RS20222162-1, RS20171777-1, RS20171884-1, RS0431-13B-1, 20152764-1, RS0431-02-1.
Why was this product recalled?
Potential for missing pins at tip of inserter.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Premia Spine, Ltd. take?
On 07/16/2025, Premia Spine Ltd. notified Premia Spine Inc. via phone call. It was requested to quarantine any suspected reworked instruments that are at the warehouse and to initiate a field action and return the affected units of the TOPS Inserter that underwent rework. No customers (hospitals/physicians) were asked to quarantine or return units as all affected units were within Premia Spine's control, however one surgeon was notified, on about 07/31/2025, of the issue and was requested to review the surgical procedure, analyze relevant patient data, including x-ray, and perform additional actions if needed to confirm the pin has not remained in the patient's body.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2516-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology