Brand Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Product Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Recall
Brand Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Product Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Model/Catalog Number: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Software Version: Not Applicable Product Description: Sterile Cover Component: Lumicell Direct Visualization System (DVS) / PMA # P230014
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Product Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2519-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 31, 2025
- FDA report date
- Sep 10, 2025
- Quantity in commerce
- 200 units (20 packs containing 10 units per pack)
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Brand Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Product Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Model/Catalog Number: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Software Version: Not Applicable Product Description: Sterile Cover Component: Lumicell Direct Visualization System (DVS) / PMA # P230014
Affected model & UDI checker
Search the exact identifiers published for Z-2519-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 900-00218 | Brand Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Product Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly | Check source notice | Not specified |
FDA code information
Model/Catalog Number: 900-00218 UDI-DIs: Sterile Cover 10-Pack Shipper Box Assembly: (01)0 0860008 53781 2, 840-00204 Assy, Dispenser Carton Sterile Cover: (01)0 0860003 25251 2, 810-00210 Assy, Outer Tray Sterile Cover: (01) 0 0860003 2520 5, Lot numbers: 43930, 43931, 43932, 43933, 42756, 44349, 44349-01, 44349-02, 44349-03, 44349-04;
Why was this product recalled?
The potential of the elastic bands included in the kit detaching or fracturing during surgical procedures. As a result, these failures may pose a risk of retained foreign bodies in the surgical cavity and potentially leading to surgical site infections.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Lumicell, Inc. take?
On July 31, 2025, URGENT: MEDICAL DECICE RECALL letters were sent to customers. Actions to be Taken by Customer: " Lumicell recommends that users dispose of the elastic bands and refrain from using them during the procedure. The Lumicell DVS may be used without elastic bands. " The elimination of elastic bands does not impact system performance. However, in the absence of elastic bands, there is a possibility that the secondary Sterile Probe Cover may shift during use. If this occurs, the Sterile Probe Cover may be repositioned, and Lumicell DVS use may continue. " If the included elastic bands are used, follow the Lumicell Direct Visualization System (DVS) Instructions for Use to ensure the bands are counted before and after the procedure to confirm the elastic bands are accounted for. " Please forward and post this notice for awareness to those who use it. If device was distributed or sold, please forward this notice to those accounts. " Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax. Please acknowledge receipt of this letter by completing the included response card. Next Actions: Lumicell is investigating the root cause of the issue and is working on a long-term solution. Lumicell will contact you in the future regarding updates. If you have any questions or need further assistance, please do not hesitate to reach out to our customer support team at 1-833-4-LUM-DVS (Monday through Friday, 9AM - 5PM Eastern Time) or email [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2519-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology