Da Vinci ASSY, PSS, SP1098, Part Number: 380601 Recall

Da Vinci ASSY, PSS, SP1098, Part Number: 380601

FDA recall Class II

Ongoing

FDA lists this Class II recall for Da Vinci ASSY, PSS, SP1098, Part Number: 380601. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2521-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 22, 2026
FDA report date
Jul 1, 2026
Quantity in commerce
454
Distribution
See key details
FDA cause
Process control

What is being recalled?

Da Vinci ASSY, PSS, SP1098, Part Number: 380601

Affected model & UDI checker

Search the exact identifiers published for Z-2521-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedDa Vinci ASSY, PSS, SP1098, Part Number: 380601Check source notice00886874114605
FDA code information

UDI: 00886874114605. Serial Numbers: 605408, 606848, 612328, 628399, 609353, 614073, 617794, 616769, 10008062, 635659, 631694, 638771, 641430, 694998, 645433, 643209, 954097, 656203, 764644, 656319, 946098, 667190, 666480, 655562, 664662, 680530, 679266, 683974, 674171, 676962, 682694, 686777, 701248, 696877, 688725, 688225, 697922, 700316, 698277, 702699, 709536, 701751, 717804, 705366, 704731, 724178, 703152, 722803, 693262, 736468, 736338, 740359, 742366, 732233, 696089, 724835, 716520, 709987, 728391, 751529, 731765, 672743, 766366, 711517, 735003, 735252, 746457, 746623, 718940, 721722, 691190, 744094, 750174, 738718, 10211920, 759253, 761448, 752536, 812651, 754935, 790597, 811147, 793019, 761000, 812650, 919988, 919987, 913011, 913012, 926423, 759791, 926422, 938794, 771653, 938795, 766365, 963841, 797872, 769736, 794437, 970568, 970562, 977045, 10003059, 957781, 973824, 804588, 10009706, 936269, 10003509, 930240, 936268, 800142, 10003510, 982782, 967667, 10009707, 973355, 944585, 10065786, 935214, 10020729, 10012963, 10020705, 10031145, 10031144, 973825, 10012962, 10036960, 963856, 10711066, 10035476, 795038, 10043635, 944583, 10072852, 10072853, 966079, 10069899, 10076490, 10087446, 10054478, 10080929, 954096, 10130089, 10165287, 942811, 10180162, 10172589, 10103483, 10103484, 10056518, 10092007, 10246370, 10209961, 10183740, 10232634, 504007, 10231116, 10211880, 10194911, 10245457, 10112470, 10189443, 977044, 10238314, 10644376, 10036961, 10274101, 10112471, 10238315, 10172574, 10281524, 10278366, 10252751, 10289629, 10293276, 10261900, 10274102, 10285085, 10281525, 10243796, 10267796, 10278365, 10285084, 10308532, 10318311, 10293274, 10323177, 10322512, 10289630, 10326939, 10293275, 10333333, 10326938, 10492299, 10302474, 10302473, 10368392, 10332310, 10346946, 10354749, 10346943, 10333332, 10346942, 10267795, 10323178, 10295600, 10223509, 10252780, 10396330, 10263730, 10354748, 10337879, 10265272, 10354750, 10384455, 10359960, 10368393, 10366206, 10375031, 10363897, 10396329, 10363898, 10375032, 10375033, 10409934, 10401560, 10409933, 10408268, 10294693, 10408269, 10396331, 10480883, 10499186, 10499185, 10384456, 10502598, 10495654, 10502599, 10505838, 10512168, 10512167, 10512166, 10508290, 10368391, 10556776, 10417803, 10526733, 10417164, 10508289, 10384454, 10492298, 10480878, 10526735, 10526736, 10537191, 10495655, 10526734, 10526732, 530114, 10537202, 10568118, 10308533, 10547839, 10564979, 10550219, 10564978, 10550220, 10550221, 10561240, 10594744, 10575228, 10571615, 10575229, 10583123, 10571614, 10556769, 10583124, 10594746, 10556777, 10564977, 10594743, 10579531, 10613173, 10568117, 10571616, 10615325, 10583126, 10619701, 10594742, 10541679, 10579530, 10561239, 10594745, 10615326, 10590323, 10590322, 10583125, 10615941, 10608089, 10644375, 10604955, 10561241, 10624252, 10647203, 10670710, 10672825, 10624990, 10295599, 10633489, 10622883, 10647201, 10624251, 10686903, 10612973, 10619702, 10672839, 10696000, 10615940, 10659191, 10651445, 10651443, 10682596, 10702363, 10644374, 10653815, 10651444, 10627633, 10686902, 10672840, 10627632, 10695999, 10665125, 10604956, 10480877, 10682597, 10694981, 10633487, 10716881, 10659189, 10728578, 10627631, 10702364, 10734551, 10659190, 10633488, 10734552, 10706206, 650202, 10744259, 10706207, 10766233, 10766236, 10793380, 10759158, 10759159, 10759161, 10763608, 10763609, 10808841, 10766234, 10763611, 10741377, 10759160, 10763610, 10826493, 10776325, 10766235, 10741378, 10744258, 10734553, 10808839, 10808840, 10771126, 10803470, 10793379, 10817400, 10798922, 10803484, 10766232, 10887260, 10830719, 10771125, 10776323, 10815457, 10744257, 10823000, 10776324, 10782092, 10873054, 10879012, 10897582, 10856718, 10889592, 10864698, 10922820, 10912617, 10647204, 10869722, 10889594, 10871773, 10845441, 10859690, 10826495, 10579532, 10782090, 10826494, 10792831, 10883203, 10871774, 10856716, 10856732, 10859689, 10947466, 10798915, 10968849, 10952495, 10957321, 10875232, 10903586, 10927370, 10947464, 10951328, 10981839, 10787962, 10908841, 10908842, 10817401, 10938311, 10933815, 10938312, 10955031, 10927371, 10933816, 10998396, 10927372, 10240354, 10998395, 11004904, 10979952, 10875231, 10912618, 10897581, 10938310, 10991256, 10991257, 10991259, 10895278, 10883202, 10889593, 10912619, 11016764, 11013103

View the original FDA recall notice

Why was this product recalled?

Operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a structural column failure, the E-brake pin would prevent E-brake engagement and may result in uncontrolled motion. In addition, if it moves/backs out it may cause restricted column range of motion and/or uncontrolled motion, which may cause patient tissue injury/crushing injury/bleeding.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Intuitive Surgical, Inc. take?

On 4/22/2026, correction notices were mailed and emailed to customers who were asked to do the following: 1. This notice needs to be passed on to all those who need to be aware within your organization or functions where the potentially affected devices have been transferred. 2. Please retain a copy of this letter, place a copy with your affected system, and keep a copy of the acknowledgement form for your files. 3. Complete the attached Acknowledgement Form immediately and return it via fax or email to Intuitive as instructed on the form. A firm Representative will schedule a site visit to perform an inspection of the system. If you need further information or support concerning this Medical Device Correction, please contact your da Vinci SP Sales Representative or contact Intuitive Customer Service at the numbers listed below: " North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: [email protected]. " Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or [email protected] " South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or [email protected] " Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or [email protected] " India: +1-800-103-6952 (9 am to 6 pm IT) " Taiwan: +0800-86-8181 (9 am to 6 pm CT)

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.