Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2522-2026

Philips Avalon Fetal Monitor, FM20, Part number M2702A, Part M2703A Recall

Philips Avalon Fetal Monitor, FM20, Part number M2702A, Part M2703A

FDA recall Class II

Ongoing

FDA lists this Class II recall for Philips Avalon Fetal Monitor, FM20, Part number M2702A, Part M2703A. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2522-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 26, 2026
FDA report date
Jul 1, 2026
Quantity in commerce
285 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Philips Avalon Fetal Monitor, FM20, Part number M2702A, Part M2703A

Affected model & UDI checker

Search the exact identifiers published for Z-2522-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedPhilips Avalon Fetal Monitor, FM20, Part number M2702A, Part M2703ACheck source notice00884838000407
FDA code information

UDI/DI 00884838000407, Serial Numbers: DE79569306, DE79569332, DE79569360, DE79569417, DE79569442, DE79569482, DE79569307, DE79569470, DE79569390, DE79569356, DE79569452, DE79569289, DE79569302, DE79569536, DE79569358, DE79569343, DE79569457, DE79569353, DE79569373, DE79569411, DE79569491, DE79569615, DE79569287, DE79569338, DE79569437, DE79569459, DE79569473, DE79569340, DE79569308, DE79569377, DE79569286, DE79569304, DE79569327, DE79569330, DE79569347, DE79569352, DE79569400, DE79569418, DE79569431, DE79569436, DE79569456, DE79569469, DE79569474, DE79569588, DE79569326, DE79569424, DE79569487, DE79569547, DE79569290, DE79569258, DE79569367, DE79569370, DE79569380, DE79569388, DE79569311, DE79569440, DE79569460, DE79569527, DE79569462, DE79569299, DE79569261, DE79569405, DE79569260, DE79569313, DE79569369, DE79569216.

View the original FDA recall notice

Why was this product recalled?

Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or permanent loss of speaker output.

What actions did Philips North America Llc take?

Philips issued an URGENT Medical Device Correction notice to its consignees on 5/26/2026 via USPS certified mail. The notice explained the issue and how it can be identified, potential risk to the patient should the loss of audio is not detected, and requested the following: "Actions that should be taken by the customer / user in order to prevent risks for patients or users " Ensure fetal heart rate tones are enabled on the Avalon Fetal Monitor until the planned fix/correction is implemented. " Pass this notice to all those who need to be aware within your organization or to any organization where affected product(s) have been potentially transferred. " Place this Urgent Medical Device Correction with the documentation of the Philips Avalon Fetal Monitor and associated devices in a place where it is most likely to be seen and viewed."

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology