Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2526-2026

Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310 Recall

Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310

FDA recall Class II

Ongoing

FDA lists this Class II recall for Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2526-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 21, 2026
FDA report date
Jul 1, 2026
Quantity in commerce
32
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310

Affected model & UDI checker

Search the exact identifiers published for Z-2526-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
147305Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310Check source noticeNot specified
147310Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310Check source noticeNot specified
FDA code information

REF/UDI-DI/Lot(Expiration): 147305/04042301061829/1049868(10/14/2028); 147310/04042301033567/1049960(10/20/28)

View the original FDA recall notice

Why was this product recalled?

A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requirements for in-vitro-cytotoxicity, which if in contact with a patient's vasculature could result in localized inflammation and tissue irritation.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did PFM MEDICAL INC. take?

On 4/21/2026, recall notices were emailed to customers who were asked to do the following: 1. Inform all customers (hospitals and clinics) and device users (physicians) who have received or used affected batches. This notification should also be shared with any organizations where the potentially affected devices have been transferred. 2. Return unused devices from the affected batches to PFM Medical, Inc. 3. Complete the enclosed Recall Response Form and email back to [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology