Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2529-2025

Artis Pheno. Image-Intensified Flouroscopic X-Ray System. Recall

Artis Pheno. Image-Intensified Flouroscopic X-Ray System.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Artis Pheno. Image-Intensified Flouroscopic X-Ray System.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2529-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 12, 2025
FDA report date
Sep 17, 2025
Quantity in commerce
20 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

Artis Pheno. Image-Intensified Flouroscopic X-Ray System.

Affected model & UDI checker

Search the exact identifiers published for Z-2529-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
10849000Artis Pheno. Image-Intensified Flouroscopic X-Ray System.Check source noticeNot specified
Not specifiedArtis Pheno. Image-Intensified Flouroscopic X-Ray System.Check source notice04056869046877
FDA code information

Model No. 10849000; UDI: 04056869046877; Serial No. 164184, 164183, 164071, 164072, 164027, 164026, 164094, 164019, 164161, 164152, 164069, 164086, 164084, 164158, 164188, 164105, 164080, 164011, 164100, 164097.

View the original FDA recall notice

Why was this product recalled?

Limited system movements after startup .

What actions did Siemens Medical Solutions USA, Inc. take?

Consignees were mailed an URGENT: MEDICAL DEVICE CORRECTION notification dated 8/12/25. The notification informs consignees that the recall issue will be corrected via release of an Update Instruction AX032/25/S by appointment, which can be arranged by contacting 1-800-888-7436. Consignees are to disseminate the recall notification to all users of affected devices within their organization and to where product may be further distributed. Consignees are to return the completed recall response form via email to [email protected] or by certified mail.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology