Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2546-2025

Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER HEAD Recall

Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER HEAD Model/Catalog Number: 804-06-311 Product Description: Material: S.S/Nitronic, Non-Sterile Orthopedic instrument for shoulder prosthesis

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER HEAD. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2546-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 18, 2025
FDA report date
Sep 17, 2025
Quantity in commerce
927 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER HEAD Model/Catalog Number: 804-06-311 Product Description: Material: S.S/Nitronic, Non-Sterile

Affected model & UDI checker

Search the exact identifiers published for Z-2546-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
804-06-311Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER HEADCheck source noticeNot specified
Not specifiedBrand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER HEADCheck source notice00190446843825
FDA code information

Lot Code: Item Number: All Lots GTIN: 00190446843825

View the original FDA recall notice

Why was this product recalled?

The reamer may kick or bind up during or immediately prior to use.

What actions did Encore Medical, LP take?

On July 29, 2025, Enovis issued a "Urgent Medical Device Recall Notification via Email to affected consignees. On August 4, 2025, Enovis issued an update to the recall notification via E-Mail requesting consignees rework affected trays and returned affected products. Enovis asked consignees to take the following actions: 1. Continue using the wedge reamer assembly while carefully inspecting and following the specific details in the supplied Quality Bulletin. 2. This notice needs to be passed on to all those within your organization or to any organization where the product has been transferred. 3. Please see the attached quality Bulletin 4. By 9/13/25, the firm will reach out regarding reworking and returning trays. 5. Acknowledge receipt of notification.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.