Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2554-2025

Reef TA Inserter, Part Number: TA2-001003, part of the Reef Interbody System Recall

Reef TA Inserter, Part Number: TA2-001003, part of the Reef Interbody System

FDA recall Class II

Ongoing

FDA lists this Class II recall for Reef TA Inserter, Part Number: TA2-001003, part of the Reef Interbody System. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2554-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 6, 2025
FDA report date
Sep 17, 2025
Quantity in commerce
74
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Reef TA Inserter, Part Number: TA2-001003, part of the Reef Interbody System

Affected model & UDI checker

Search the exact identifiers published for Z-2554-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedReef TA Inserter, Part Number: TA2-001003, part of the Reef Interbody SystemCheck source notice10889981290733
Not specifiedReef TA Inserter, Part Number: TA2-001003, part of the Reef Interbody SystemCheck source notice10889981437800
FDA code information

UDI-DI: 10889981290733, 10889981437800. Lots: BT107376C, BT112289C, BT112779C, BT107376CR1, BT107376CR2, BT107376CR1R2, BT112289CR2, BT112779CR2

View the original FDA recall notice

Why was this product recalled?

Inserter, used as part of a system with an intervertebral body fusion device, may contribute to difficulty disengaging the implant or improper implant positioning, which may may lead to prolonged surgery, intraoperative revision, and may require the use of alternate instrumentation. Devices may also have incorrect UDI-DI marking.

What actions did SeaSpine Orthopedics Corporation take?

On 8/6/2025, recall notices were emailed to customers and distributors who were informed of the following: A laser marking error was identified in which the TA2-001003 inserters were incorrectly marked with the UDI-DI for model TA2-001002 (10889981290733). Updated Revision E Inserters have been shipped to you via overnight delivery. Kindly replace all existing Revision C/R1/R2 Inserters with the enclosed Revision E units. Return the Revision C Inserters to the recalling firm using provided RGA and prepaid shipping label. This recall notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where potentially affected devices have been transferred. Should you have any questions regarding this matter, please contact the firm at: [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology