Thuvera Laser Console. Software Version 2.2. Portable Thulium Fiber Laser. Recall

Thuvera Laser Console. Software Version 2.2. Portable Thulium Fiber Laser.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Thuvera Laser Console. Software Version 2.2. Portable Thulium Fiber Laser.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2554-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 5, 2026
FDA report date
Jul 1, 2026
Quantity in commerce
198 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Thuvera Laser Console. Software Version 2.2. Portable Thulium Fiber Laser.

Affected model & UDI checker

Search the exact identifiers published for Z-2554-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
MD2300F-001507Thuvera Laser Console. Software Version 2.2. Portable Thulium Fiber Laser.Check source noticeNot specified
Not specifiedThuvera Laser Console. Software Version 2.2. Portable Thulium Fiber Laser.Check source notice00810071230226
FDA code information

Model No. MD2300F-001507; UDI-DI 00810071230226; Serial No. MDUF250342, MDUF250400, MDUF250429, MDUF250060, MDUF250061, MDUF250067, MDUF250068, MDUF250069, MDUF250071, MDUF250076, MDUF250083, MDUF250086, MDUF250091, MDUF250092, MDUF250104, MDUF250131, MDUF250133, MDUF250134, MDUF250159, MDUF250160, MDUF250199, MDUF250200, MDUF250202, MDUF250203, MDUF250204, MDUF250205, MDUF250240, MDUF250241, MDUF250242, MDUF250278, MDUF250297, MDUF250299, MDUF250312, MDUF250314, MDUF250325, MDUF250340, MDUF250341, MDUF250351, MDUF250352, MDUF250355, MDUF250367, MDUF250368, MDUF250370, MDUF250380, MDUF250384, MDUF250385, MDUF250395, MDUF250399, MDUF250409, MDUF250410, MDUF250427, MDUF250430, MDUF250431, MDUF250433, MDUF260009, MDUF260010, MDUF260011, MDUF260017, MDUF260019, MDUF260023, MDUF260024, MDUF260025, MDUF260026, MDUF260027, MDUF260032, MDUF260043, MDUF260067, MDUF260068, MDUF260070, MDUF260072, MDUF260077, MDUF260079, MDUF260094, MDUF260095, MDUF260096, MDUF260097, MDUF260098, MDUF260099, MDUF260100, MDUF260101, MDUF260103, MDUF260104, MDUF260105, MDUF260106, MDUF260107, MDUF260108, MDUF260109, MDUF260112, MDUF260115, MDUF250014, MDUF250015, MDUF250016, MDUF250050, MDUF250051, MDUF250052, MDUF250053, MDUF250054, MDUF250055, MDUF250056, MDUF250057, MDUF250058, MDUF250059, MDUF250062, MDUF250063, MDUF250064, MDUF250065, MDUF250066, MDUF250070, MDUF250072, MDUF250073, MDUF250074, MDUF250075, MDUF250084, MDUF250085, MDUF250088, MDUF250089, MDUF250090, MDUF250093, MDUF250102, MDUF250103, MDUF250105, MDUF250132, MDUF250145, MDUF250158, MDUF250161, MDUF250162, MDUF250198, MDUF250201, MDUF250206, MDUF250207, MDUF250220, MDUF250237, MDUF250238, MDUF250239, MDUF250243, MDUF250245, MDUF250279, MDUF250280, MDUF250295, MDUF250296, MDUF250298, MDUF250310, MDUF250311, MDUF250313, MDUF250323, MDUF250344, MDUF250348, MDUF250349, MDUF250350, MDUF250353, MDUF250357, MDUF250364, MDUF250365, MDUF250369, MDUF250371, MDUF250372, MDUF250378, MDUF250379, MDUF250381, MDUF250382, MDUF250383, MDUF250386, MDUF250393, MDUF250396, MDUF250397, MDUF250398, MDUF250407, MDUF250408, MDUF250411, MDUF250412, MDUF250428, MDUF250432, MDUF260016, MDUF260018, MDUF260020, MDUF260031, MDUF260041, MDUF260044, MDUF260069, MDUF260071, MDUF260078, MDUF260088, MDUF260089, MDUF260090, MDUF260091, MDUF260092, MDUF260093, MDUF250087, MDUF250208, MDUF250244, MDUF250324, MDUF250326, MDUF250354, MDUF250356, MDUF250394, MDUF260042, MDUF260045, MDUF260076; Software Version 2.2.

View the original FDA recall notice

Why was this product recalled?

Software anomaly that causes a false display of error code 5018.

What actions did IPG Medical Corporation take?

An URGENT MEDICAL DEVICE RECALL notice dated 05/05/2026 was emailed to IPG Medical's consignee notifying them of this recall correction to address Error 5018 and Error 5007/5023 displayed on monitors with software version 2.2. The notification provides instructions to consignees to work around the error messages until a software upgrade is deployed at a later date. The notification instructs consignees that their staff may correct units onsite via software upgrade or units may be returned to IPG Medical for a factory correction when the update is available. A corrected IFU will be provided to all customers. Questions can be directed via phone to 774-633-5903 or [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology