Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2566-2026

Surgify Halo, 5.4 mm, Short Recall

Surgify Halo, 5.4 mm, Short, Model/Catalog Number: 54.070.NVG.H1; drills, burrs, trephines & accessories (simple, powered)

FDA recall Class II

Ongoing

FDA lists this Class II recall for Surgify Halo, 5.4 mm, Short. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2566-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Apr 30, 2026
FDA report date
Jul 1, 2026
Quantity in commerce
N/A
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Surgify Halo, 5.4 mm, Short, Model/Catalog Number: 54.070.NVG.H1; drills, burrs, trephines & accessories (simple, powered)

Affected model & UDI checker

Search the exact identifiers published for Z-2566-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
54.070.NVG.H1Surgify Halo, 5.4 mm, ShortCheck source noticeNot specified
Not specifiedSurgify Halo, 5.4 mm, ShortCheck source notice06429811532090
FDA code information

UDI: 06429811532090 ; All lots until the IFU update has been implemented

View the original FDA recall notice

Why was this product recalled?

Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS).

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did SURGIFY MEDICAL OY take?

Surgify issued a Medical Device Correction Notice - Surgify Halo - Restriction of Use notice to its consignees on 4/30/2026 via email. The notice explained the issue, potential risk, and requested the following: "What steps must the user take? Please review your current practice and ensure the device is only used in accordance with the updated safety information. Instruction to Customers: The customers have been instructed to do the following: - Do not use this device in endoscopic surgical procedures. If endoscopic access is required, an alternative instrument should be selected. Specifically: . Do not use this device in endoscopic surgical procedures. If endoscopic access is required, an alternative instrument should be selected. . Ensure that all relevant clinical staff and surgical team members are made aware of this notice. . Please retain this notice for your records and attach it to the relevant product documentation. Please acknowledge receipt of this Medical Device Correction Notice by replying to [email protected] within 30 business days, confirming that the notice has been read, understood, and communicated to all relevant users within your organization. Please include your facility name, your name, and your role. If you have previously used the device in endoscopic procedures without incident, no patient follow-up is required; however, this application should now be discontinued in line with the updated safety information." For questions regarding this matter, please contact us by phone at +358 105 176 310 or via email at [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology