Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-2573-2025

UniCel DxH 690T COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C34520 Recall

UniCel DxH 690T COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C34520

FDA recall Class II

Ongoing

FDA lists this Class II recall for UniCel DxH 690T COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C34520. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2573-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 13, 2025
FDA report date
Sep 24, 2025
Quantity in commerce
N/A
Distribution
See key details
FDA cause
Component change control

What is being recalled?

UniCel DxH 690T COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C34520

Affected model & UDI checker

Search the exact identifiers published for Z-2573-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
C34520UniCel DxH 690T COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C34520Check source noticeNot specified
Not specifiedUniCel DxH 690T COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C34520Check source notice15099590729110
FDA code information

UDI: 15099590729110/ Serial Numbers: BD17734, BE12048, BE28105, BF14045, BF23096, BF46236, BG26135, BG50247, BH12019, BH43168

View the original FDA recall notice

Why was this product recalled?

Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did BECKMAN COULTER, INC. take?

On August 13, 2025, Beckman Coulter initiated a "Urgent Medical Device Recall" notification via E-Mail. Beckman Coulter expanded the recall to additional consignees September 25, 2025 Beckman Coulter asked consignees to take the following actions: For specimens with a WBC count > 30 x 103 and = 90 x 103 cells/L, follow the actions listed within the customer letter. For specimens with a WBC count > 90 x 103 cells/L, follow the actions listed within the customer letter. To identify WBC counts > 30 x 103 cells/L, consider implementing decision rules within your Laboratory Information System (LIS), Middleware, or on affected instruments. Consult your Medical Director to determine if a retrospective review of results is warranted. Beckman Coulter recommends posting this letter on or near the affected analyzers Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected products listed above to another laboratory, please provide them a copy of this letter. Acknowledge receipt of the notification.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.