Heartstring III Proximal Seal System, 3.8 MM. Intravascular anastomosis occluder. Recall
Heartstring III Proximal Seal System, 3.8 MM. Intravascular anastomosis occluder.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Heartstring III Proximal Seal System, 3.8 MM. Intravascular anastomosis occluder.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-2587-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 15, 2025
- FDA report date
- Sep 24, 2025
- Quantity in commerce
- 28,331 units (16,615 US, 11,716 OUS)
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Heartstring III Proximal Seal System, 3.8 MM. Intravascular anastomosis occluder.
Affected model & UDI checker
Search the exact identifiers published for Z-2587-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| HSK-3038 | Heartstring III Proximal Seal System, 3.8 MM. Intravascular anastomosis occluder. | Check source notice | Not specified |
FDA code information
Model No. HSK-3038; UDI: 00607567700314¿; Serial No. 3000377764, 3000378278, 3000379222, 3000379987, 3000383569, 3000388598, 3000390377, 3000394076, 3000396875, 3000398149, 3000400694, 3000402932, 3000403566, 3000404026, 3000404996, 3000405699, 3000407693, 3000409172, 3000409619, 3000410091, 3000410583, 3000412000, 3000412963, 3000415599, 3000416141, 3000417121, 3000417711, 3000418572, 3000418772, 3000419866, 3000420311, 3000422320, 3000422757, 3000423694, 3000424888, 3000425725, 3000428927, 3000429846, 3000431259, 3000431934, 3000434091, 3000434862, 3000435356, 3000436032, 3000436446, 3000440202, 3000441040, 3000441162, 3000442134, 3000443538, 3000443910, 3000444136, 3000446374, 3000447343, 3000448483, 3000449214, 3000449897, 3000450561, 3000451914, 3000452532, 3000453432, 3000453804, 3000455728, 3000456346, 3000457969, 3000458434, 3000459087, 3000459889, 3000460034, 3000461587, 3000464463, 3000465229, 3000465676, 3000465679, 3000466111, 3000466878, 3000467559, 3000470706, 3000471492, 3000473348, 3000473350, 3000475032, 3000475033, 3000476158, 3000476611, 3000477574, 3000478410, 3000479294, 3000479465, 3000481661, 3000484284, 3000485833, 3000488966, 3000377764, 3000378278, 3000379222, 3000379987, 3000388598, 3000391049, 3000394076, 3000396875, 3000398149, 3000400694, 3000403566, 3000404026, 3000404996, 3000405699, 3000409172, 3000409619, 3000410091, 3000412000, 3000412963, 3000415599, 3000417121, 3000417752, 3000418572, 3000419866, 3000420311, 3000422320, 3000423694, 3000424888, 3000425725, 3000428927, 3000431259, 3000431934, 3000434091, 3000434862, 3000435356, 3000436032, 3000436109, 3000436446, 3000437664, 3000440202, 3000440757, 3000441162, 3000442134, 3000442635, 3000443538, 3000443910, 3000444136, 3000446374, 3000447140, 3000447343, 3000448483, 3000449214, 3000449897, 3000450292, 3000451914, 3000452532, 3000452651, 3000453432, 3000453804, 3000455728, 3000456197, 3000456346, 3000457780, 3000457969, 3000458434, 3000459889, 3000460034, 3000461587, 3000464463, 3000465228, 3000465229, 3000465676, 3000465679, 3000466111, 3000466878, 3000467559, 3000470706, 3000471492, 3000473348, 3000475032, 3000475033, 3000476158, 3000477574, 3000478410, 3000479294, 3000479465, 3000481661, 3000484284, 3000484896, 3000488349, 3000488742, 3000488966.
Why was this product recalled?
Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
What actions did Maquet Cardiovascular, LLC take?
Consignees were sent an URGENT MEDICAL DEVICE CORRECTION notification dated August 2025. The notification provides instructions for consignees to carry out in the instance of failure of the Heartstring Seal to load, failure of the Heartsring III Seal to deploy into the aortotomy, and failure of the deployed Heartstring III Seal to provide adequate hemostasis. Consignees are asked to forward the recall notification to individuals within their organization who use the recalled devices and to organizations to which product was further distributed. Consignees are to return the completed response form to [email protected]. Questions about this recall can be directed to Getinge Customer Service at 1-888-880-2874. ****Update 4/17/2026**** On August 15, 2025 URGENT MEDICAL DEVICE CORRECTION lupdate letters were sent to customers providing updated instructions for consignees to carry out in the instance of failure of the Heartstring Seal to load, failure of the Heartsring III Seal to deploy into the aortotomy, and failure of the deployed Heartstring III Seal to provide adequate hemostasis. Consignees are asked to forward the recall notification to individuals within their organization who use the recalled devices and to organizations to which product was further distributed. Consignees are to return the completed response form to [email protected]. Questions about this recall can be directed to Getinge Customer Service at 1-888-880-2874.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2587-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology