Recall at a glance
- Recall number
- Z-2590-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 17, 2025
- FDA report date
- Sep 24, 2025
- Quantity in commerce
- 4,300
- Distribution
- See key details
- FDA cause
- Labeling mix-ups
What is being recalled?
SureStep Foley Tray System Lubri-Sil I.C. Complete Care Infection Control Foley Catheter Tray, REF: A303414A
Affected model & UDI checker
Search the exact identifiers published for Z-2590-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| A303414A | SureStep Foley Tray System Lubri-Sil I.C. Complete Care Infection Control Foley Catheter Tray, REF: A303414A | Check source notice | Not specified |
FDA code information
Lot# NGJY3685
Why was this product recalled?
Foley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with infection control coating, when it's silicone hydrogel with antimicrobial coating, or 3)Natural latex with infection control, when it's silicone hydrogel coated without antimicrobial coating, which may cause infection and/or local and/or systemic allergic reaction
What actions did C.R. Bard, Inc. take?
On 7/17/2025, recall notices were mailed and emailed to Recall Coordinator, Director of Nursing, Director of Purchasing, Director of Risk Management, and Distributors who were asked to do the following: 1) It is recommended that clinicians discontinue the use of impacted devices and find alternative catheters with the appropriate characteristics required for their patients to continue therapy and avoid potential health risks related to this issue. 2) Quarantine and discard all devices within your facility s control per your facility s procedures. 3) Share this notice with any users of the product within your facilities to ensure they are also aware of this Medical Device Recall (Correction). 4) If you have purchased this product through a distributor, please contact this distributor to request appropriate credit or order alternate replacement product. 5) Complete and return the attached Customer Response Form via email to [email protected] In addition, distributors were asked to: a) Identify all customers within your distribution network that purchased any affected product as defined in this notification. Provide a copy of the attached customer letter to all customers to advise them of this field action notification on the recalling firm's behalf. b) Contact recalling firm if you require assistance with credits to be issued for discard-ed products or for replacement product options. If you require assistance contact the firm: Medical Information Services: 1-800-555-7422, [email protected]; Technical Support: 1-844-823-5433, [email protected]
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-2590-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology