Sterile Medline Convenience Kits: 1) EXTREMITY PACK Recall

Sterile Medline Convenience Kits: 1) EXTREMITY PACK, Model Number: DYNJ45701B

FDA recall Class I

Ongoing

FDA lists this Class I recall for Sterile Medline Convenience Kits: 1) EXTREMITY PACK. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-2593-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Aug 8, 2025
FDA report date
Oct 8, 2025
Quantity in commerce
88 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Sterile Medline Convenience Kits: 1) EXTREMITY PACK, Model Number: DYNJ45701B

Affected model & UDI checker

Search the exact identifiers published for Z-2593-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
DYNJ45701BSterile Medline Convenience Kits: 1) EXTREMITY PACKCheck source noticeNot specified
Not specifiedSterile Medline Convenience Kits: 1) EXTREMITY PACKCheck source notice10193489856279
Not specifiedSterile Medline Convenience Kits: 1) EXTREMITY PACKCheck source notice40193489856270
FDA code information

1) DYNJ45701B, UDI-DI: 10193489856279(each), 40193489856270(case), Lot Number: 25GBF686

View the original FDA recall notice

Why was this product recalled?

Convenience kits labeled as sterile have not gone through the sterilization process.

What actions did Medline Industries, LP take?

Customers were alerted via email and telephone beginning 8/8/25 of a potential issue with the products. An URGENT MEDICAL DEVICE RECALL notification letter dated 8/22/25 was sent to customers. This communication is a follow-up to the notification you received from your Medline Sales Representative and to acknowledge the subsequent return of impacted product to Medline. This communication is to provide formal documentation of the recall for your records. No additional action is required to be taken. Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Sterile Medline Convenience Kits. It has been identified that specific convenience kits labeled as sterile have not gone through the sterilization process. As a result, these kits are non-sterile. If the non-sterile kits are assumed sterile upon use, this could result in potential contamination of the sterile-field and an increased risk of patient infection that may require the use of intravenous or oral antibiotics. Refer to the enclosed list for all affected item number(s) and affected lot number(s). Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax and to Medline Industries, LP. If you have any questions regarding this notification, please contact the Director, Regulatory Compliance at ([email protected]). or Product Recall Specialist at [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.